3-349-086-03
14/12.14
GMC-I Messtechnik GmbH
METRISO C
Insulation and Resistance Measuring Instrument
Special features for insulation resistance measurement
Quick testing with limit value and signal lamp
Auto-ranging for insulation resistance measurement over the
entire scale range for quick determination of the measured
insulation value
Special features for resistance measurement (low-resistance)
Quick testing with limit value and signal lamp
Automatic polarity reversal for recognition
of interference voltage
Special instrument features
Hold function: the measured value is frozen at the display
after the measurement key is released.
Measured values can be stored to memory with reference
to electrical circuits, distribution cabinets and other objects
thanks to alphanumeric entry.
Data interface for transmission of measured values, and
for software updates
Convenient report generating software,
can be expanded to a comprehensive database
Display
The LCD window consists of a backlit dot matrix which is used
to display menus, configuration options and measurement
results, as well as online help. Various user interface languages
can be selected, depending upon the country in which the test
instrument is used.
Operation
The instrument is very easy to operate. A multifunction key allows
for one-handed operation when selecting menus and starting
measurements. Basic functions and sub-functions are selected
with the help of four softkeys.
Battery Charge Level Indicator and Device Self-Test
A battery symbol in the main menu with 5 segments ranging from
depleted to fully charged keeps the user continuously informed
concerning battery charge level.
The test instrument is switched off automatically if the batteries
are depleted, and it includes a charge control circuit for safe
charging of rechargeable NiMH or NiCd batteries.
Test patterns can be queried one after the other in the self-test
mode, and display LEDs and relays can be tested.
Sturdy Housing for Rugged Use
Soft plastic jacketing protects the instrument against impacts, or
if it is inadvertently dropped.
Battery powered insulation resistance measuring instrument
in accordance with DIN VDE 0413, parts 2 and 4
for measurement in systems with nominal voltages of up to 500 V
Insulation resistance measurement
Measuring range 0 100 G
Variable test voltage: adjustable from 50 to 1000 V
Low-resistance measurement
Measuring range 0 100
Temperature and humidity measurement
via IrDa interface with additional adapter
Contact current measurement
Measuring range 0 10 mA
METRISO C
Insulation and Resistance Measuring Instrument
2 GMC-I Messtechnik GmbH
Signal Lamps
The device recognizes errors in the electrical system automati-
cally, which are indicated with four lamps, (see following table).
Data Interface
Measurement data can be read out to a printer or a PC via the
integrated IRDA interface, providing the user with 3 advantages.
Transmission of stored data to a PC for processing and
archiving, or for the generation of official reports
Immediate print-out of all measurement data (via adapter)
Software Updates
The test instrument will never become obsolete thanks to
software updates which can be installed via the IRDA interface.
Updates can be performed by our service department as part of
our re-calibration service, or by the user himself.
Applicable Regulations and Standards
Sample Displays
Lamp Status Function
red
Potential difference between finger contact and
measurement input is greater than 150 V
Netz
Mains
blinks red
Mains voltage is present at the measurement
inputs, insulation resistance measurement is
disabled
LIMIT red
Measured insulation resistance is below the
allowable limit value.
Measured low resistance has exceeded the
allowable limit value.
Measured contact current has exceeded the
allowable limit value.
U>25V red
A voltage of greater than 25 V is present at the
measurement inputs.
Discharging in not yet complete.
IEC 61010-1/ EN 61010-1
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
Part 1: General requirements
(IEC 61010-1:2010 + Cor. :2011);
German edition EN 61010-1:2010
IEC 61557
DIN EN 61557
Part 1:2007
Part 2:2007
Part 4:2007
Part 10:2001
VDE 0413
Part 1:2007
Part 2:2008
Part 4:2007
Part 10:2001
Electrical safety in low voltage distribution systems up to
1000 V a.c. and 1500 V d.c. – Equipment for testing, mea-
suring or monitoring of protective measures
Part 1: General requirements
Part 2: Insulation resistance
Part 4:2007 Resistance of earth connection
and equipotential bonding
Part
10:2001 Combined measuring equipment for
testing, measuring or monitoring of
protective measures
IEC 61326-1:2005
DIN EN 61326-1:2006
VDE 0843-20-1:2006
Electrical equipment for measurement, control and
laboratory use – EMC requirements -
Part 1: General
requirements
DIN EN 60529
VDE 0470-1:2000
Test instruments and test procedures,
protection provided by enclosures (IP code)
DIN EN 1081:1998 Testing floor coverings in explosive atmospheres
for electrostatic discharge capacity
Main Menus
Insulation Resistance Measurement
Voltage Measurement
Low-Resistance Measurement
Contact Current Measurement Online Help
Temperature and Humidity
Measurement