2
System Capabilities
• Supports simultaneously:
- Two inlets
- Three detectors (third detector
as TCD or ECD)
- Four detector signals
• State-of-the-art detector electronics
and the full-range digital data path
enable peaks to be quantified over
the entire concentration range of the
detector (10
7
for the FID) in a single
run.
• Full EPC is available for all inlets and
detectors. Control range and resolu-
tion are optimized for the specific
inlet or detector module.
• Up to six EPC modules can be
installed, providing control of up to
16 channels of EPC.
• Pressure setpoint and contol preci-
sion to 0.001 psi provides more
retention time locking precision for
low-pressure applications.
• EPC with capillary columns provides
four column flow control modes:
constant pressure, ramped pressure
(3 ramps), constant flow, or ramped
flow (3 ramps). Column average
linear velocity is calculated.
• Atmospheric pressure and tempera-
ture compensation is standard, so
results do not change, even when
the laboratory environment does.
• Low Thermal Mass (LTM) Series II
system can be added to achieve
fastest cycle times via rapid LTM
capillary column module heating and
cooling.
• Serial port interface for Remote
Advisor or Optional Bar Code Reader.
• One-button access to maintenance
and service modes from the
keyboard
• Preprogrammed leak tests
• Automatic Liquid Sampling is fully
integrated into mainframe control.
• Setpoint and automation control can
be done from the local keyboard or
via a networked data system. Clock-
time programming can be initiated
from the front panel to initiate
events (on/off, method start, etc.) at
a future date and time.
• A run time deviation log is created
for each analysis to ensure that all
method parameters were achieved
and maintained.
• A full array of traditional gas sam-
pling and column switching valves
are available
• 550 timed events
• Display of all GC and ALS setpoints
at the GC or data system.
• Context-sensitive online help
Column Oven
• Dimensions: 28 × 31 × 16 cm.
Accommodates up to two 105 m ×
0.530 mm id capillary columns or two
10-ft glass packed columns (9 in. coil
diameter, 1/4 in. od), or two 20-ft
stainless steel packed columns
(1/8 in. od).
• Operating temperature range suit-
able for all columns and chromato-
graphic separations. Ambient
temperature +4 °C to 450 °C.
- With LN
2
cryogenic cooling:
–80 to 450 °C.
- With CO
2
cryogenic cooling:
–40 to 450 °C.
• Temperature setpoint resolution:
0.1 °C.
• Supports 20 oven ramps with
21 plateaus. Negative ramps are
allowed.
• Maximum achievable temperature
ramp rate: 120 °C/min (120 V units
are limited to 75 °C/min, see
Table 1).
• Maximum run time: 999.99 min
(16.7 h).
• Oven cool down (22 °C ambient) 450
to 50 °C in 4.0 min (3.5 min with
oven insert accessory).
• Ambient rejection: < 0.01 °C per 1 °C.
Electronic Pneumatics
Control (EPC)
• Compensation for barometric pres-
sure and ambient temperature
changes is standard.
• Pressure has typical control of
±0.001 psi for the range of 0 to
150 psi. Pressure setpoints may be
adjusted in increments of 0.001 for
the range 0.000 to 99.999 psi;
0.01 psi for the range 100.00 to
150.00 psi.
• User may select pressure units as
psi, kPa, or bar.
• Pressure/flow ramps: Three
maximum.
• Carrier and makeup gas settings
selectable for He, H
2
, N
2
, and
argon/methane.
• Flow or pressure setpoints for each
inlet or detector parameter with both
Agilent 7890B and Agilent
ChemStations.
Table 1. Typical 7890B GC Oven Ramp Rates
Temperature 120 V Oven* Fast ramp rates** (°C/min)
range (°C) rates (°C/min) Dual-Channel Single-Channel***
50 to 70 75 120 120
70 to 115 45 95 120
115 to 175 40 65 110
175 to 300 30 45 80
300 to 450 20 35 65
* Results obtained with line voltage maintained at 120V
** Fast ramp rates require power > 200 volts at > 15 Amps.
***Requires G2646-60500 oven insert accessory.