Luna 6415
Lightwave Component Analyzer
The Luna 6415 Lightwave Component Analyzer is a fast and simple-
to-use tool for testing passive optical components and modules.
The 6415 measures and analyzes the Insertion Loss (IL) and Return
Loss (RL) distribution, as well as length, working in either reection
or transmission mode.
The 6415 utilizes optical frequency domain reectometry (OFDR)
technology to measure backscattered or transmitted light as a
function of distance. Extremely high sensitivity and sampling
resolution (20 μm) make the 6415 an ideal analyzer for photonic
integrated circuits (PICs) and silicon photonics. The Luna 6415
reduces the cost and complexity of test while increasing throughput
by measuring RL, IL and length in reection or transmission with a
single instrument.
KEY FEATURES
APPLICATIONS
Return loss (RL) and insertion loss
(IL) analysis
Analyze components in reection
and transmission
Trace distributed RL over length
of optical path
Spectral analysis of RL and IL
Detect and precisely locate
reective events and measure
path length (up to 200 m)
• Speed, resolution and accuracy
for optimizing production test
- 20 μm sampling resolution
• Spatial RL testing
Automated IL test and analysis
• Skew measurement with sub-
picosecond resolution
PLCs, waveguide devices, AWGs,
ROADMs, etc.
Couplers, switches, beam splitters
High-Speed and High-Resolution
OFDR Measurements for
Manufacturing Test
Measuring in reflection mode, the Luna 6415 measures return loss versus length. The
bottom plot shows the spectral content of the identified reflection event (filter).
PERFORMANCE
PARAMETER SPECIFICATION
Measurement
Sampling resolution (two-point)
1
20 µm
Measurement length modes
Reection 20, 50, 200
2
Transmission 40, 100, 400
2
Wavelength accuracy
3
1.5 pm
Time-of-ight delay accuracy
3
± 0.0034 %
Center wavelength 1546.69 nm
Measurement time (20 m length mode) 0.08 s
Maximum optical power 5 mW
Return Loss Measurements (Reection Mode)
RL dynamic range
4
70 dB
Total range
5
0 to -130 dB
Sensitivity
5
-135 dB
Resolution
6
± 0.1 dB
Accuracy
6
± 0.5 dB
Insertion Loss Measurements (Reection/Transmission)
IL dynamic range, in transmission mode 70 dB
IL dynamic range, in reection mode
7
15 dB
Resolution
8
± 0.1 dB
Accuracy
8
± 0.2 dB
Physical and Interfaces
Remote interface SCPI API over TCP/IP
Optical connector type FC/APC
Operating power (max) 50 W
Weight (controller not included) 13 lb (6 kg)
Case size
13.5 x 13.9 x 4.3 in (34 x 35 x 11 cm)
6415 REV.5 06.01.21
NOTES
1. Distance between two sample points along the length axis in SMF-28.
2. With extended range option OPT06450.
3. Accuracy guaranteed via internal NIST-traceable HCN gas cell.
4. Range between strongest reflection greater than -60 dB and noise floor.
5. Noise floor return loss at half of maximum length.
6. Measured with 1 cm integration width.
7. Two way loss before backscatter reaches noise floor and IL
measurements are no longer possible.
8. Measured with 10 cm integration width.
Luna 6415 Lightwave Component Analyzer
+1.866.586.2682
solutions@lunainc.com
www.lunainc.com
*Specifications subject to change without notice.
Reection Measurements
Total Insertion Loss (IL)
Spectral amplitude response
Total path length
Reectivity, RL versus length
Event loss measurement (RL, IL)
RL spectral amplitude analysis
• Event length measurement
Transmission Measurements
ORDERING
Product # Description Includes
Luna 6415
Lightwave
Component Analyzer
Luna 6415 mainframe for C band with
measurement length modes of 20 m and
50 m, instrument controller (workstation-
class laptop), application software and
accessory kit.
OPT06450
Extended range
option
Extends measurement length of Luna
6415 to 200 m in reflection (400 m in