Lightwave Component Analyzer
The Luna 6415 is a fast and simple-to-use analyzer for passive
optical components and modules. The Luna 6415 measures and
analyzes the Insertion Loss (IL) and Return Loss (RL) distribution, as
well as length, working in either reection or transmission.
The Luna 6415 utilizes optical frequency domain reectometry
(OFDR) technology to measure backscattered or transmitted light
as a function of distance. Extremely high sensitivity and sampling
resolution (20 μm) make the Luna 6415 an ideal testing tool for
photonic integrated circuits (PICs) and silicon photonics. The Luna
6415 reduces the cost and complexity of test while increasing
throughput by measuring RL, IL and length in reection or
transmission with a single instrument.
KEY FEATURES
APPLICATIONS
Return loss (RL) and insertion loss
(IL) analysis
Trace distributed RL over length of
optical path
Spectral analysis of RL and IL
Detect and precisely locate
reective events and measure
path length
• Speed, resolution and accuracy
for optimizing production test
- 20 μm sampling resolution
- 20 m measurement range
- 6 Hz scan/acquisition rate
• Spatial RL testing
Automated IL test and analysis
• Skew measurement with sub-
picosecond resolution
PLCs, waveguide devices, AWGs,
ROADMs, etc.
Couplers, switches, beam splitters
High-Speed and High-Resolution
OFDR Measurements for
Manufacturing Test
Measuring in reflection mode, the Luna 6415 measures return loss versus length. The
bottom plot shows the spectral content of the identified reflection event (grating).
PERFORMANCE
PARAMETER SPECIFICATION UNITS
Measurement
Sampling resolution (two-point)
1
20 µm
Time-of-ight delay accuracy
2
± 0.0034 %
Maximum device length
Reection 20 m
Transmission 40 m
Wavelength range 40 nm
Wavelength accuracy
2
1.5 pm
Center wavelength 1546.69 nm
Measurement rate 6 Hz
Maximum optical power 5 mW
Return Loss Characteristics (Reection Mode)
RL dynamic range
3
70 dB
Total range
4
0 to -130 dB
Sensitivity
4
-135 dB
Resolution
5
± 0.1
dB
Accuracy
5
± 0.5
dB
Insertion Loss Characteristics (Reection/Transmission)
IL dynamic range, in transmission mode 70 dB
IL dynamic range, in reection mode
6
15 dB
Resolution
7
± 0.1
dB
Accuracy
7
± 0.2
dB
Physical
Optical connector type FC/APC
Operating power (max) 50 W
Weight (controller not included) 13 (6) lb (kg)
6415 REV.1 04.30.19
NOTES
1. Distance between two sample points along the length axis in SMF-28.
2. Accuracy guaranteed via internal NIST-traceable HCN gas cell.
3. Range between strongest reflection greater than -60 dB and noise floor.
4. Noise floor return loss at half of maximum length.
5. Measured with 1 cm integration width.
6. Two way loss before backscatter reaches noise floor and IL
measurements are no longer possible.
7. Measured with 10 cm integration width.
Luna 6415 Lightwave Component Analyzer
+1.866.586.2682
solutions@lunainc.com
www.lunainc.com
*Specifications subject to change without notice.
Reection Measurements
Total Insertion Loss (IL)
Spectral amplitude response
Total path length
Reectivity, RL versus length
Event loss measurement (RL, IL)
RL spectral amplitude analysis
• Event length measurement
Transmission Measurements
ORDERING
Product # Description Includes
Luna 6415
Lightwave
Component Analyzer
Luna 6415 mainframe for C band,
application software, instrument
controller (workstation-class laptop)