GMC-I Messtechnik GmbH
METRISO PRIME+
Digital High-Voltage Insulation Tester
German
Accreditation Body
D-K-15080-01-01
Features
Test Voltages to 5000 V
The instrument is suitable for non-destructive measurement of
insulation resistance in electrical systems, as well as in machines,
transformers, cables and electrical equipment utilized in, for
example, locomotives, street cars and ocean going vessels with
selectable test voltages of up to 5 kV.
Voltage Measurement to 1000 V
Testing for absence of voltage at the device under test in systems
of up to 1 kV can be performed with the voltage measuring range.
Discharging Capacitive Devices Under Test
Capacitive devices under test such as cables and coils, which
may be charged by the test voltage, are discharged by the mea-
suring instrument. The falling voltage value can be observed at
the display.
Measurements per EN 61557 Parts 1 and 2 (VDE 0413)
Nominal current amounts to 1 mA at a test voltage of 100 V,
250 V, 500 V or 1000 V.
Highly Insulated Measurement Cables
The highly insulated measurement cables are permanently con-
nected for safety reasons, and due to technical measuring con-
siderations. Danger resulting from inadvertently disconnected
cables, for example in the event of charging caused by capacitive
devices under test, is thus avoided.
Polarization Index
A polarization index test is recommended for electrical machines.
This procedure involves expanded testing of insulation resistance.
DC measuring voltage from the METRISO PRIME+ is applied to
the insulation for a duration of 10 minutes. Measured values are
documented after one minute, and after ten minutes. If the insula-
tion is good, the value measured after ten minutes is higher than
the value measured after one minute. The relationship between
the two measurement values is the polarization index. Charged
material within the insulation is aligned due to the application of
measuring voltage over a long period of time, resulting in polariza-
tion. The polarization index indicates whether or not the charged
material contained in the insulation can still be moved, thus allow-
ing for polarization. This, in turn, is an indication of the condition of
the insulation.
Data Management and Report Generation
The data of each measurement can be stored under a selected
object number. Furthermore, a description for this object can be
entered via the keyboard of the optional PSI module (Feature I1).
The data management function allows for individual measurement
data of a previously selected object to be displayed and to be
deleted if required, or for previously entered objects to be deleted.
Depending on the number of stored objects (max. 254), up to
1,600 measurements can be stored. The current memory occu-
pancy is continuously displayed as a bar graph.
Report data can be printed out at an external printer with Cen-
tronics interface via PSI module (Feature I1) or via printer adapter
DA-II (accessory).
Furthermore, it is possible to create report templates at a PC
which can be downloaded to the test instrument.
Large measuring range from 0.4 M ... 1 T
Variable test voltages, or in fixed steps of
100 V, 250 V, 500 V, 1.0 kV, 1.5 kV, 2.0 kV, 2.5 kV, 5.0 kV
Polarization index and absorption ratio
Voltage measurements to 1000 V
Frequency measurement from 15 Hz to 1 kHz
Capacitance measurement from 0.1 to 5 F
Measurement of electrical discharge
Guard terminal for the compensation of surface currents
5 m extension cable included as accessory
Supply power from mains, internal set of storage batteries or
external
12 V
supply
Backlit dot matrix display
Digital display of measured values and limit values
Timer function: 1 second to 100 minutes
Data logger function
DAkkS calibration certificate
Applications
Insulation measurement in large systems, and for cables, motors,
generators etc.
METRISO PRIME+
Digital High-Voltage Insulation Tester
2 GMC-I Messtechnik GmbH
Characteristic Values
Measuring Ranges:
Standard DIN EN 61557-1:2007
DIN EN 61557-2:2008
VDE Regulation VDE 0413-1:2007
VDE 0413-2:2008
Insulation Resistance
Test duration: automatic (until measured value is stable),
manual (1 to 120 s) or continuous measurement (lock function)
Polarization Index (PI), Absorption Ratio (DAR)
PI and DAR are calculated values. The specifications of the insulation measurement
are applicable.
Insulation Test Voltage
Variable test voltages are adjustable in increments of 50 V
Short-circuit current up to 1.00 kV, test voltage 2mA
Voltage Measurement
Frequency Measurement
Voltage of measuring quantity: 50 V ... 1 kV
Breakdown Voltage
Capacitance Measurement
Dielectric Discharge (DD)
Reference Conditions
Ambient
temperature +23 C 2K
Relative humidity 40 ... 60%
Measured quantity
frequency 50 Hz 10 Hz (during voltage
measurement)
Line voltage
waveshape Sinusoidal, deviation between RMS and
rectified value < 1%
Power Supply METRISO PRIME+
Line voltage 207 V 253 V / 49 Hz 61 Hz
or (depending on country-specific version)
Feature A43:
108 V
132 V / 59 Hz 61 Hz
Power consumption < 18 VA
Storage batteries NiMH 9.6 V, 3 Ah, charging period 6 hours
Number of
measurements
at nominal current
as per VDE 0413 700
Ambient Conditions
Accuracy 0 C ... + 40 C
Operating temperature 5 C ... + 40 C
Storage temperature 20 C ... + 60 C (without batteries)
Relative humidity max. 75%, no condensation allowed
Elevation to 2000 m
Deployment indoors, outdoors: only in the specified
abient conditions
Display
Range []
Measuring
Range
Test Voltage Intrinsic
Uncertainty
Measuring
Uncertainty
0.00 M ... 50.0 G 0.60 M ...10.0 G 100 V ... 250 V ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
> 10.0 G ... 50.0 G ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
0.00 M ... 250 G
0.40 M ... 50.0 G > 250 V ... 1.00 kV ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
> 50.0 G ... 250 G ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
0.00 M ... 999 G
0.40 M ... 200 G >1.00 kV ... 5.00kV ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
> 200 G ... 999 G ±(7% rdg. + 6 d) ±(10% rdg. + 8 d)
t1 [min] t2 [min] Limit [min]
PI
00:00 ... 01:00 ... 99:50 00:00 ... 10:00 ... 99:50 0.10 ... 4.00 ... 9.80
DAR
00:00 ... 00:30 ... 99:50 00:00 ... 01:00 ... 99:50 0.10 ... 1.60 ... 9.80
Nominal Values
of Test Voltage
Variable
Test Voltage
Nominal
Current
Intrinsic
Uncertainty
100 V, 250 V, 500 V, 1.00 kV
1.0 mA 0 ... +25% rdg.
1.50 kV, 2.00 kV, 2.50 kV
0.4 mA ± 5% rdg.
5.00 kV
0.1 mA ± 3.5% rdg.
100 V...1.00 kV 1.0 mA ± 15% rdg.
> 1.00 kV...2.50 kV 0.4 mA ± 5% rdg.
> 2.50 kV...5.00 kV 0.1 mA ± 3.5% rdg.
Measuring range
Frequency
[Hz]
Impe-
dance
Intrinsic
Uncertainty
Measuring
Uncertainty
test voltage
dc
50 V ... 5.00 kV
——
(2.5%
rdg.
+ 5 d)
(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc
15 ... 500
1 M
(2.5%
rdg.
+ 2 d)
(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc
>500...1 k
1 M
(10%
rdg.
+ 2 d)
(12.5%
rdg.
+ 5 d)
Measuring Range Impedance Intrinsic
Uncertainty
Measuring
Uncertainty
15.0 Hz 1.00 kHz 1 M
(0.5% rdg. + 2 d)
(1% rdg. + 2 d)
Parameters Setting Range Intrinsic
Uncertainty
Measuring
Uncertainty
Voltage range 100 5000 V
(10% rdg. + 8 d) (15% rdg. + 10 d)
Rise time 5 300 s
Measuring time 1 120 s / auto /
cont. measurement
——
Display
Range
Measuring
Range
Test
Voltage
Intrinsic
Uncertainty
Measuring
Uncertainty
0.00
10.0
F 0.10
5.00
F
100
450 V
(10% rdg. + 5 d) (15% rdg. + 8 d)
5005 kV
(5% rdg. + 5 d) (10% rdg. + 8 d)
Limit
DD 0.10 ... 2.00 ... 9.80