3-349-923-03
5/1.19
GMC-I Messtechnik GmbH
METRISO PRIME10
High-Precision Insulation, Low-Resistance and Voltage Meter
Applications
Continuous display of measured insulation resistance or
leakage current
Automatic discharging of the device under test at the end of
insulation testing
Acoustic signal at 5 second intervals for quick generation of a
time-resistance characteristic
Adjustable measuring times of up to 99'59"
Test times T1, T2 and T3 for measuring one or two absorption
coefficients within a range of 1 to 600 s
Polarization index (PI), absorption coefficients Ab1 and Ab2,
dielectric absorption ratio (DAR)
Display of momentarily applied test voltage during measurement
Test current: 1.2 mA, 3 mA and 5 mA
Insulation testing by means of 2 or 3-wire method
Measurement can be conducted with measurement cables
with lengths of up to 55 meters
Automatic measurement of multi-core cables with the optional
AutoISO-5000 test adapter (max. voltage: 5 kV)
Capacitance measurement during RINS insulation test
Measurement of temperature with probe as accessory
Dielectric discharge (DD)
Fault localization by means of pulse control mode
Adjustable limit values for measured resistance values for R
INS
und R
CONT
Measurement of leakage current during insulation test
Direct and alternating voltage measurement from 0 to 750 V
Graphic representation of insulation resistance at the display
during measurement
New memory structure with storage of measuring points,
systems and customer data
Use of a miniature Bluetooth keyboard (optional)
5.6" LCD display with background illumination
Keyboard illumination
Mains operation or with rechargeable lithium-ion battery
Internal quick charger
Applicable Regulations and Standards
IEC 61010-1/EN 61010-1/
VDE 0411-1
Safety requirements for electrical equipment for mea-
surement, control and laboratory use
– General requirements
DIN EN 61557 / VDE0413 Part 1:2007-12 General requirements
Part 2:2008-02 Insulation resistance measuring
instruments
Part 4:2007-12 Resistance of earth connection and
equipotential bonding
Part 10: 2014-03
Combined measuring equipment
for testing, measuring or
monitoring protective measures
EN 60529
VDE 0470, part 1
Test instruments and test procedures
Degrees of protection provided by enclosures (IP code)
DIN EN 61326-1
VDE 0843-20-1
Electrical equipment for measurement, control and labo-
ratory use – EMC requirements – Part 1: General re-
quirements
Insulation measurement per EN 61557-2/VDE 0413, part 2
Test voltage in fixed steps:
50 V, 100 V, 250 V, 500 V, 1000 V, 2500 V, 5000 V, 10,000 V
Measurement with incrementally rising voltage
Measuring range up to 40 TΩ per IEC 61557-2
Measurement of polarization index and absorption ratio
Measurement with shielded measurement cable
Protection against voltage conducting objects
Variable adjustment of limit values
Digital filter for stabilizing measured values
Creation of R/I or R/U diagrams
Storage of test results
Low-resistance measurement per EN 61557-4/VDE 0413,
part 4, continuity testing of protective conductors and equipo-
tential bonding connections with a test current of > 200 mA
Bluetooth and USB interface for transferring data to a PC
METRISO PRIME10
High-Precision Insulation, Low-Resistance and Voltage Meter
2 GMC-I Messtechnik GmbH
Characteristic Values
Alternating/Direct Voltage Measurement
Frequency range: 45
... 65 Hz
Insulation Resistance Measurement
Test voltage accuracy (R
obc
[Ω] 1000 • U
N
[V]):
-0, +10% of the selected value
Measuring range per IEC 61557-2:
U
N
= 10,000 V: 10.0 MΩ ... 40.0 TΩ
Measurement with Rising Alternating Voltage with AutoISO-5000
(U
INS
5 kV)
Intrinsic uncertainty can be calculated for all other voltages using
the following formula:
δ
R
= ±(3%+(UINS/(U
INS
-R
zm
•21•10
-12
)-1) • 100%) ±10 digits
Where:
U
INS
= selected test voltage [V]
R
zm
= measured resistance [Ω]
Maximum values for measured resistance depend on the selected
test voltage. See following list:
Note: No degree of accuracy is specified for the R
INSOmin
mea-
surement because the test instrument conducts this measure-
ment with a selectable test current. This results in the following
calculation:
Where:
R
INSmin
= minimum insulation resistance
measured without current limiting
U
INSnom
= nominal test voltage
I
INSnom
= nominal test current (1.2, 3 or 5 mA)
Additional intrinsic uncertainty of 3-wire measurement (caused
by “G” connection): 0.05% caused by reduced leakage cur-
rent via 250 kΩ resistor with a measurement via 100 MΩ and
a test voltage of 50 V
Max. short-circuit current: 6 mA ±15%
Remaining charge at objects depends on test current:
1.2, 3, 5 mA
Measurements with AutoISO-5000
Leakage Current Measurement
* The measurement’s resolution and electrical unit of measure result from the mea-
suring range and the individual insulation resistance value.
** Calculation is based on the resistance measurement.
Capacitance Measurement
Capacitance measurements are conducted during RISO mea-
surements (while the device under test is being discharged).
Intrinsic uncertainty of the measurement corresponds to a
measured capacitance value and a resistance of greater than
10 MΩ connected in parallel.
No measuring error has been defined for measuring voltages of less
than 100 V.
Cable length L is calculated as C/Cx, and intrinsic uncertainty
depends on the measuring range.
Time constant TC is calculated as Rins • C, and intrinsic
uncertainty depends on the measuring range.
Displayed Range Resolution Intrinsic Uncertainty
0.0 V ... 29.9 V 0.1 V ±(2% rdg. + 20 digits)
30.0 V ... 299.9 V 0.1 V ±(2% rdg. + 6 digits)
300 V ... 750 V 1 V ±(2% rdg. + 2 digits)
Displayed Range Resolution Intrinsic Uncertainty
000 kΩ ... 999 kΩ 1 kΩ
±(3% rdg. + 10 digits)
1.00 MΩ ... 9.99 MΩ 0.01 MΩ
10.0 MΩ ... 9.9 MΩ 0.1 MΩ
100 MΩ ... 999 MΩ 1 MΩ
1.00 GΩ ... 9.99 GΩ 0.01 GΩ
10.0 GΩ ... 99.9 GΩ 0.1 GΩ
100 GΩ ... 999 GΩ 1 GΩ±(3.5% rdg. + 10 digits)
1.00 TΩ ... 9.99 TΩ 0.01 TΩ±(7.5% rdg. + 10 digits)
10.0 TΩ ... 40.0 TΩ
where U
N
= 10 kV 0.1 TΩ±(12.5% rdg. + 10 digits)
Voltage Measuring
Range
Measuring Range
for AutoISO-5000
50 V 200 GΩ 20.0 GΩ
100 V 400 GΩ 40.0 GΩ
250 V 1.00 TΩ 100 GΩ
500 V 2.00 TΩ 200 GΩ
1000 V 4.00 TΩ 400 GΩ
2500 V 10.0 TΩ 400 GΩ
5000 V 20.0 TΩ 400 GΩ
10000 V 40.0 TΩ
Displayed Range Resolution Intrinsic Uncertainty
000 kΩ ... 999 kΩ 1kΩ±(3% rdg. + 10 digits) due to the
tester ± 1% additional uncer-
tainty due to the AutoISO-5000
1.00 MΩ ... 9.99 MΩ 0.01 MΩ
10.0 MΩ ... 99.9 MΩ 0.1 MΩ
100 MΩ ... 999 MΩ 1 MΩ
1.00 GΩ ... 9.99 GΩ 0.01 GΩ
10.0 GΩ ... 99.9 GΩ 0.1 GΩ
100 GΩ ... up to the value at
which additional uncertainty of
the AutoISO-5000 amounts to
5%
1 GΩ±(3% rdg. + 10 digits) due to the
tester ± 5% additional uncer-
tainty due to the AutoISO-5000
Displayed Range Resolution Intrinsic Uncertainty
0 ... 1.2mA
***0 ... 3mA
0 ... 5mA
Displayed Range Resolution Intrinsic Uncertainty
0 nF ... 999 nF 1 nF
±(5% rdg. + 5 digits)
1.00 μF ... 49.99 μF 0.01 μF
ISOnom
nomISO
ISO
I
U
R =
min