Innovating Together
Rigel Medical, Bracken Hill, South West Industrial Estate,
Peterlee, County Durham, SR8 2SW United Kingdom
Tel: +44 (0) 191 587 8730 Fax: +44 (0) 191 586 0227
Email: enquiry@rigelmedical.com Web: rigelmedical.com
Part of
Need Help?
For technical issues please visit rigelmedical.com and view the Application
Notes in the Downloads tab. Application Notes are regularly updated and will
most likely cover any common issues.
For further assistance please contact your Rigel Medical supplier or use the
following contact details to speak to a member of the Rigel Medical team:
Sales and Delivery enquiries
Tel: +44 (0) 191 587 8730 Fax:+44 (0) 191 586 0227
Email: sales@rigelmedical.com
Technical enquiries
Tel: +44 (0) 191 587 8701
Email: support@rigelmedical.com
Service, Calibration and Repair
Tel: +44 (0) 191 587 8739 Fax: +44 (0) 191 518 4666
Email: service@calibrationhouse.com
Remember to activate
your 2 year warranty*
by registering your Uni-Therm with Rigel Medical
Go to rigelmedical.com/registerproduct
*Terms and Conditions apply
Rev 1.3
Part No. 398A554
Quick Start Guide
Rigel Uni-Therm
Electrosurgical Analyser
Introduction
The Rigel Uni-Therm electrosurgical analyser is a versatile bench-mounted yet portable
tool for testing the performance of electrosurgical equipment.
Utilising a large, full colour graphical display, the Rigel Uni-Therm is the first fully
stand-alone electrosurgical analyser that can offer true intuitive, user friendly
operation and operator safety.
What’s in the box?
RUN POWER TEST GRAPH VIEW
GOTO
DATA
STOP
HOLD
1
4
If any items are missing on receipt of your new unit, please contact your equipment supplier
immediately.
User notes
5
3
2
i
h
The following symbols are used throughout this Rigel Uni-Therm Quick Start Guide.
Indicate[s] a hazardous situation which, if not avoided, could result in death or serious injury.
Indicate[s] a hazardous situation which, if not avoided, could result in minor or moderate injury.
CAUTION [signs] may be used to alert against unsafe practices that can result in
equipment damage.
Warnings and Cautions
Users - The Rigel Uni-Therm electrosurgical analyser is designed for use by adequately trained
technical personnel only.
Operation - The Rigel Uni-Therm electrosurgical analyser is designed for use within the published
specifications. Any application outside of these specifications or any unauthorised user
modifications may result in hazardous conditions or improper operation.
Operating conditions - When operating the Rigel Uni-Therm at its maximum capability, during
certain power tests and in an ambient temperature of 40°C, it is possible for the surface
temperature of the top cover to exceed 50°C but not exceed 7C as is permitted under IEC 61010
design requirements.
Operating conditions - The Rigel Uni-Therm is designed to operate between 15°C and 40°C.
Exposure to temperatures outside of this range may adversely affect the performance of the device.
Connections - Do not connect active power electrodes to the front panel of the Rigel Uni-Therm.
The front panel is for low voltage connections only (ie. REM connection/ footswitch). Failure to
comply with this may lead to internal damage of components and will void the warranty.
Cleaning - The Rigel Uni-Therm case can be cleaned with a damp cloth with, if necessary, a small
amount of mild detergent. Do not allow liquid inside the Rigel Uni-Therm or near the sockets.
Do not use abrasives, solvents, or alcohol.
h
h
h
i
i
i
1 Rigel
Uni-Therm
electrosurgical analyser
2 IEC Mains Power Lead
3 USB cable
4 4mm test link (x3)
5 Calibration certificate
1
RUN POWER TEST GRAPH VIEW
GOTO
DATA
STOP
HOLD
Large colour graphic display
On/Off Button
Function Keys F1 - F4
Rotary encoder for menu navigation
Start/OK Button
Stop/End Button
Front connection panel
Folding legs at front base of unit
Front connection panel
Getting to know your Rigel Uni-Therm
Connect the REM/plate security
electrodes to the front panel for the REM test.
Connect the footswitch contacts to the front panel for automatic footswitch control.
Side connection panel
Connect the HF active and neutral electrodes to
the side panel using the on-screen connection
diagrams for reference.
The scope output can be used in conjunction with the additional scope cable
accessory (P/N: 367A950) to view the waveform shape.
Rear panel
Ensure the ventilation holes are free from obstruction during
operation to allow for maximum unrestricted air flow.
Power fuses are located below the IEC connector.
2
Before you switch on!
Ensure the correct incoming mains supply voltage range is selected
The mains selector switch is located on the base of the unit. Please ensure this switch is
set to either 230V AC ± 10% or 120V AC ± 10% as required.
Ensure legs are deployed
The Rigel Uni-Therm has two legs at the front base of the unit. These are designed to raise
the front of the unit to allow clear airflow for the base fan and also to improve the viewing
angle of the colour graphic display.
Ensure all ventilation holes are free from obstruction
The Rigel Uni-Therm has several fan and air vents located around the casing.
Please ensure these are free from obstruction to allow for maximum air flow.
Active power connections
Ensure no active power electrodes are connected to the front panel of the Rigel Uni-
Therm. The front panel is for low voltage connections only (i.e. CQM connection or
footswitch)
Instruction Manual
This
Quick Start Guide is designed to be used
in conjunction with the full Rigel Uni-Therm
instruction manual.
The Rigel Uni-Therm electrosurgical analyser
instruction manual is available as a download
from rigelmedical.com/rigel-downloads
The download section also includes a free trial
for our asset management software,
Med-eBase.
3
Getting Started
In this section we will identify the Rigel Uni-Therm electrosurgical analyser’s basic functions
in order to get you started.
From the setup menu select
Time/Date
Use the rotary encoder to
change the Time/Date and
press function key F4 to Save
Change the language settings
Set the time/date
Select Setup
Select Language
Scroll to the required language
and select to confirm
From the start screen use the
rotary encoder or function key
F1 to select Menu
4
Perform a test
REM Test
Select the REM test from
the main menu;
Select automatic or manual
control and press the green
button to start the test
(for automatic mode only)
To stop the test, press
CAPTURE ALARM.
HF Leakage Test
Select the HF LEAKAGE
test from the main menu;
Select the appropriate test
using the rotary encoder.
Change between the graph
mode and description mode,
using the function keys.
Power Test
Select the POWER test from
the main menu;
Select either CONTINUOUS
or GRAPH mode.
Ensure that the settings (i.e.
CUT or COAG) correspond
with the connections as
these control the automatic
footswitch control.
Connections - Remember to
use the on-screen connection
diagrams as a reference for
connecting the active and
neutral electrodes and test
links.
i
The Uni-Therm interface has been segmented into a safe, low voltage area (front panel)
and high voltage area (side panel)
Switch on
Turn on your Rigel Uni-Therm by pressing and holding the
green ON button until the Rigel splash screen appears.