V14
1
CFM150M SERIES
150W MEDICAL AND ITE OPEN FRAME
AC-DC MODULES
WITH PFC
Features
* Universal Input Range 90 ~ 264VAC
* Medical and ITE Safeties
* 3”x 5” Compact Size
* Less than 1 U high
1.05”
* Industry Standard Pin Out
* Active PFC Meets EN61000-3-2
* High Efficiency up to 93%
* Meets CISPR/FCC Class B
* Remote Voltage Sense
* Over Voltage Protection
* Continuous Short Circuit Protection
* No load Power Consumption<0.5W
* Meets 2MOPP
Specifications are subject to change without notice.
MODEL
OUTPUT
VOLTAGE
MAX.
LOAD
MIN.
LOAD
RIPPLE &
NOISE
NOTE 1
VOLTAGE
ACCURACY
NOTE 2
LINE
REGULATION
NOTE 3
LOAD
REGULATION
NOTE 4
% EFF.
(Typ.)
NOTE 5
CFM150M120
12 V 12.5 A
0A 1% ±1% ±0.5% ±1% 90%
CFM150M240
24 V 6.25 A
0A 1% ±1% ±0.5% ±1% 92%
CFM150M360
36 V 4.17 A
0A 1% ±1% ±0.5% ±1% 92%
CFM150M480
48 V 3.13 A
0A 1% ±1% ±0.5% ±1% 93%
V14
2
Specifications
CFM150M Series Derating Curve
INPUT SPECIFICATIONS:
Voltage....…................................................... 90~264Vac
Frequency ………………………………………….……………..... 47 to 63Hz
Inrush Current ………………….……………................. 110A max. @240Vac
Leakage Current ………………………………………….…...…… 300uA max.
OUTPUT SPECIFICATIONS:
Hold-up Time ………………………………………………………... 16ms typ.
Short Circuit Protection ……….…….………… Hiccup mode(Auto Recovery)
Adjustment Range on Vout ………………………………………………. ±5%
Over Voltage Protection ………………………… Recycle AC input to restart
Temperature Coefficient ……………………..............................… ±0.05%/
GENERAL SPECIFICATIONS:
Isolation ………………………………….……… Input to output = 5,656VDC
NOTE:
Operating Temperature .......................... -20
~ 70
(see Derating Curve)
1. Add a 0.1uF ceramic capacitor and 10uF E.L.
Storage Temperature ……………….…………….........……........ -20 ~ 85
capacitor to output for Ripple & Noise measuring @20MHz BW.
Humidity ....................................................... 93% RH max. Non condensing
2. Voltage accuracy is set at 100% rated load and 25
Ta.
Cooling ……………………………………………………. Natural Convection
3. Line regulation is measured from high line to low line with full load.
Switching Frequency ………………………............................ 90KHz Typical.
4. Load regulation is measured from full to 10% load.
MTBF ………….… MIL-HDBK-217F, GB, 25
/115VAC …… 100Khrs min.
5. Typical efficiency at 230VAC and full load at 25
.
Altitude ……………………..………………...………………………….. 3000m
6. Standard input and output connectors wafer with LONG CHU P3060
Dimensions ………….. 5.000x3.000x1.050inches (127.00x76.20x26.67mm)
series and mate with MOLEX housing 09-50-1031 and 09-50-1081 or
Weight ………………………………………………………... 270g (0.6Pounds)
equivalent.
SAFETY AND EMC:
7. DC output pin 1, 2, 3, 4: Vout (-), DC output pin 5, 6, 7, 8: Vout (+).
Emission and Immunity (Ed. 4.0) …………..… EN55011, EN55032 Class B
FCC CFR 47 Part 15, EN61000-3-2, 3, IEC61000-4-2, 3, 4, 5, 6, 8, 11
Safety (Ed. 3.1) …………………………………. IEC60601-1:2005+A1:2012
IEC60950-1:2005+A1:2009+A2:2013, EN60601-1:2006+A11:2011+A1+A12
EN60950-1:2006+A11+A1+A12:2011+A2:2013, UL ANSI/AAMI ES60601-1:2005, UL60950-1, 2
nd
Editon
CFM150M Series
All Dimensions are in inches[mm]
Tolerances Inchesx.xxx=±0.02
Millimeters
x.xx=±0.5
2
3
4
5
6
7
1
8
ACL
ACN
RS-
RS+
VR
LED
0.224(5.70)
4.546(115.48)
5.000(127.00)
0.224(5.70)
2.555(64.90)
3.000(76.20)
0.118(3.00)max
0.063(1.60)
1.050(26.67)max
Typical at 25˚C, nominal line and 75% load, unless otherwise Specified
0%
20%
40%
60%
80%
100%
-20 -10 0 10 20 30 40 50 60 70
Ambient Temperature(
o
C)
Rated Load(%)
20 LFM(Natural C onve cti on)