GS 04L55B01-01EN
Data Acquisition System
GM
GM10 Data Acquisition Module
GM90MB Module Base
GM90PS Power Supply Module
OVERVIEW
The Data Acquisition System GM is a data logger
that excels in versatility and expandability. The main
unit includes on-board memory for data acquisition
and also supports SD cards for external storage. The
system consists of a Data Acquisition Module (GM10),
Power Supply Module (GM90PS), and Module Base
(GM90MB), which is used to mount a variety of
modules.
SMARTDAC+ GM utilizes the same modules as the
GX/GP series of SMARTDAC+.
High expandability and maintainability: Employs
YOKOGAWAs original block structure, making
it easy to rearrange the combination of required
modules. The structure also excels in maintainability
because even after installation, every module can be
removed separately.
Multichannel measurement: Measures up to 420
analog input channels on the GM10-2 and up to 100
channels even on the GM10-1.
Flexible system conguration: A wide variety of
systems, such as multichannel measurement from
10 to 420 channels or data acquisition through
communication,canbecongured.
High environmental tolerance: The guaranteed
operating temperature range of -20°C to 60°C allows
more freedom in where the system can be installed.
Mounting:
Not only can the system be installed on a
desktop, it can also be mounted on DIN rails or a wall.
Less wiring through distribution: Installation of sub
units away from the main unit makes measurement
possible without connecting sensor signal wires over
a long distance.
Long-term recording and saving: Large internal
memory (500 MB on the GM10-1 and 1.2 GB on the
GM10-2) allows long-term recording and saving.
Secure saving of recording data: An SD card (SD/
SDHC) up to 32 GB (1 GB included) can be used
for the external storage medium. And the FTP client
function can be used to provide data redundancy
usingaleserver.
Rich I/O interface:
Seven types of I/O modules are
available: analog input, analog output, digital input,
digital output, digital I/O, pulse input and PID control.
DCV (DC voltage), TC (thermocouple), RTD (resistance
temperature detector), DI (contact or TTL level voltage),
mA (DC current) can be assigned to each channel as
input signals.
Web server function: Various settings can be
conguredonlinefromaWebbrowser,without
using a dedicated software application. Moreover,
measured data can be monitored in real time.
●High withstand voltage performance: Safe
measurement up to withstand voltage of 600 V
between input terminal and ground is possible by
using a high withstand voltage analog input module.
●High-speed measurement: Measurement at the
shortest interval of 1 ms is possible by using a high-
speed analog input module.
●Dual interval measurement: Measurement can be
performedbysettingtwodierentscanintervals.
Simple settings: The standard USB port makes it
easy to connect with the PC.
Supports 21 CFR Part 11: GM supports the USA
FDA’s Title 21 CFR Part 11 regulation by advanced
security function (/ AS optional).
●PID control function: PID control on up to 20 loops
is possible by installing a PID control module.
●Program control function (/PG option): Program
control of up to 99 patterns is possible by using a
PID control module and program control function. In
addition, up to 32 time event and 32 PV events can
be used.
* For details on the PID control module, PID control
function, and program control function (/PG), see the
GX90UTPIDControlModuleGeneralSpecications
(GS04L53B01-31EN).
●Math function (/MT option): Various calculations
are possible. The report function enables creation of
hourly, daily, and monthly reports and other types of
data.
Logic computation outputs results of computation as
0s or 1s to internal switches and DO channels.
●Extensive network functions: The standard
Ethernet interface allows network features such as
variouse-mailnoticationsandFTPletransfer.In
addition, Modbus/TCP, Modbus/RTU (/C3), EtherNet/
IP (/E1), and CC-Link familty SLMP communication (/
E4) are supported.
Universal Viewer software: A free software
application is available for displaying and printing
waveforms of measured data on a PC.
GM10 GM90MB GM90PS
Yokogawa Electric Corporation
2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
GS 04L55B01-01EN
©Copyright October 2014
14th Edition Apr. 08, 2021
[Release 4]
Thecontentsofthisgeneralspecications
correspond to the GM with release number 4 and
style number 2.
Releasenumber:rmwareIDnumber(GM10)
Style number: hardware ID number (GM90PS)
General
Specications
Apr. 08, 2021-00
Actual values underlying accurate measurements
Input Type Measuring accuracy
*1
(typical
value
*2
)
DCV 20 mV ±(0.01%ofrdg+5μV)
60 mV ±(0.01%ofrdg+5μV)
6V (1-5V) ± (0.01% of rdg +2 mV)
TC
*3
R, S ±1.1°C
B ±1.5°C
K
(-200.0 to
1370.0°C)
± (0.01% of rdg +0.2°C for 0.0 to
1370.0°C;
± (0.15% of rdg +0.2°C) for -200.0
to 0.0°C
K
(-200.0 to
500.0°C)
±0.2°C for 0.0 to 500.0°C;
± (0.15% of rdg +0.2°C) for -200.0
to 0.0°C
J ± 0.2°C for 0.0 to 1100.0 °C;
± (0.10% of rdg + 0.2 °C) for
-200.0 to 0.0 °C
T ± 0.2°C for 0.0 to 400.0°C;
± (0.10% of rdg + 0.2 °C) for
-200.0 to 0.0 °C
N ± (0.01% of rdg + 0.2°C) for 0.0 to
1300.0 °C;
± (0.22% of rdg + 0.2°C) for -200.0
to 0.0 °C
RTD Pt100
(-200.0 to
850.0°C)
± (0.02% of rdg + 0.2°C)
Pt100
(high resolution)
(-150.00 to
150.00°C)
± (0.02% of rdg + 0.16°C )
rdg: Reading value
*1 These values have been calculated from the
accuracy testing data at the time of shipment of
the instrument from the factory.
Applies to GX90XA-10-U2, A/D integration time
16.67 ms or more, General operating conditions:
23±2 ºC, 55±10% RH, supply voltage 90–132,
180–264 V AC, power frequency within 50/60
Hz ±1%, warm-up of 30 minutes or more, no
vibrations or other hindrances to performance.
*2 For the measuring accuracy (guaranteed),
seethemodule’sgeneralspecications(GS
04L53B01-01EN).
*3 These values do not include the reference
junction compensation accuracy.
CONFIGURATION
SMARTDAC+ GM handles a wide variety of
applications through the combination of various
modules.
The unit that includes GM10 (Data Acquisition Module)
is called the main unit. A unit connected to the main
unit via GX90EX (expansion module) is called a sub
unit. Modules in a unit can be connected by installing a
GM90MB (Module Base).
Unit Types
Main Unit (Single Unit)
A unit consisting of a GM10 and a GM90PS. Up to 10
I/O modules* can be connected to a unit.
* Up to 8 when a GX90XA-T1 is connected
Main Unit (Multi Unit)
A unit consisting of a GM10, a GM90PS and a
GX90EX. Up to six I/O modules can be connected to
a unit. Up to six sub units can be connected via the
GX90EX.
Sub Unit
A unit consisting of a GM90PS and a GX90EX. Up
to six
*
I/O modules can be connected to a unit. The
main unit and sub units are connected directly using
LAN cables. The maximum connection distance
between two units is 100 m. It can also be used as an
expandable I/O for the GX/GP paperless recorder.
* See “Restrictions and Cautions” on page 26.
Module Types
Model Description
GM10 Data Acquisition Module for SMARTDAC+ GM
A module that acquires data from I/O modules
and expansion modules.
A main unit requires one module of this type.
GM90PS Power Supply Module for SMARTDAC+ GM
A module that supplies power to the modules
connected in the unit.
A unit requires one module of this type.
GX90EX* Expansion module
A module that connects units to expand the
system.
A main unit or a sub unit requires one module
of this type.
GM90MB Module Base
This is used to connect modules (excluding the
GM90PS).
* ThermwareversionofGX90EXthatcanbeusedonthe
SMARTDAC+ GM must be R1.02.01 or later.
Forthedetailedspecicationsoftheexpansion
module,seethefollowinggeneralspecications.
Material No.: GS 04L53B00-01EN
2
All Rights Reserved. Copyright © 2014, Yokogawa Electric Corporation GS 04L55B01-01EN