Arbitrary Function Generators
AFG31000 Series Datasheet
The Tektronix AFG31000 Series is a high-performance AFG with built-in
arbitrary waveform generation, real-time waveform monitoring, and the
largest touchscreen on the market. Providing advanced waveform
generation and programming capabilities, waveform verification, and a
modern touch-screen interface, the new AFG31000 is sure to delight and
simplify the job of every researcher and engineer.
Key performance specifications
1 or 2 channel models
Output amplitude range 1 mV
P-P
to 10 V
P-P
into 50 Ω loads
Basic (AFG) mode:
25 MHz, 50 MHz, 100 MHz, 150 MHz, or 250 MHz sine waveforms
250 MSa/s, 1 GSa/s or 2 GSa/s sample rates
14-bit vertical resolution
Built-in waveforms include sine, square, ramp, pulse, noise, and
other frequently used waveforms
Sweep, Burst, and Modulation modes (AM, FM, PM, FSK, and
PWM)
Advanced (Sequence) mode:
Continuous mode (optional Sequence, Triggered and Gated
modes)
16 Mpts arbitrary waveform memory on each channel (128 Mpts
optional)
Up to 256 steps in sequence mode with loop, jump and wait events
Variable sampling clock 1 µSa/s to 2 GSa/s
Key features
Patented InstaView
technology enables engineers to see the actual
waveform at the Device Under Test (DUT) in real time, without the
need of an oscilloscope and probe, eliminating the uncertainty caused
by mismatched impedance
Sequencing option adds the ability to program long, complex
waveforms with up to 256 steps
The 9-inch capacitive touch screen works like a smart phone and has
short-cuts to frequently used settings
Built-in ArbBuilder lets you create and edit arbitrary waveforms on the
instrument, eliminating the need to connect to a PC
Outputs are protected from over voltage and current to minimize
potential instrument damage
Compatible with TekBench
software to help students set up, control,
and analyze test results in the lab
Applications
Advanced research
Clock and system synchronization
Replication of real world signals
Component and circuit characterization and validation
Embedded circuit design and test
General purpose signal generation
www.tek.com 1
Basic and Advanced Modes
The AFG31000 series is the industry’s first arbitrary function generator with
full function Basic (AFG) and Advanced (Sequence) modes.
In Basic mode, the AFG31000 generates traditional functions and arbitrary
waveforms. The touchscreen and front-panel controls make it simple to set
up.
Basic mode lets you change frequency without the need to worry about
waveform length and sample rate. This feature is useful in analog designs
that characterize filter/amplifier frequency responses or in digital designs
where clock rates change frequently.
Key settings are visible at a glance, and are easy to adjust using touch, numeric keypad,
or rotary controls
New with the AFG31000, Advanced mode provides the ability to generate
multiple waveforms with complex timing. In this mode, you can compose a
list (or a sequence) of 1 to 256 waveforms, with total waveform length up to
16 Mpts/ch (128 Mpts/ch optional) and define the ouput sequence of these
waveforms. Repeat, go-to, wait, jump, and triggered events are all
supported and the large memory provides space to store many waveforms
or long waveforms.
This feature is very useful in applications where many test cases need to
be performed sequentially. Instead of loading the test cases one by one,
you can put all of them in a sequence and load at one time, switching from
one to another seamlessly to greatly improve the test efficiency.
Advanced mode lets you build complex waveform sequences with flexible step controls
Sequenced sine waveforms with different frequency and amplitude.
Additionally, Advanced mode uses variable sample rate technology. Every
sample in a waveform is output once and only once in each cycle,
synchronized to the sample rate. Since there is no skipping or repetition, all
details in the waveforms are kept. This feature is very useful for
applications in which signal fidelity is extremely critical, such as IQ
modulation and pulse train generation.
InstaView
technology shows the actual
waveform at the DUT
Most waveform generators assume they are driving a 50 Ω impedance.
However, most devices under test do not have a 50 Ω impedance. This
mismatch results in an inconsistency between the waveform as set on the
AFG and the signal at the DUT.
Datasheet
2 www.tek.com