JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Email: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 62 50 29
Email: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: +1 315 437 5866
Fax: +1 315 437 5860
Email: info.us@jumo.net
Internet: www.jumousa.com
Data Sheet 404304 Page 1/5
40430400T10Z001K000
Reference conditions
to DIN 16 086 and DIN IEC 770/5.3
Measuring ranges
see ordering details
Overload limit
ranges 0 to 400 mbar
5× full scale
ranges above 400 mbar
2× full scale
Maximum system pressure
(for p measurements)
1 bar (on each side)
Bursting pressure
for all ranges 2bar
Parts in contact with medium
Ni, Al, CuBe, PU
Output
0 to 10 V, burden 2 k
0/4 to 20 mA, burden 500
4 to 20 mA, burden (U
B
- 12 V) ÷ 0.02 A
2-wire
Burden error
0.2 %
Zero offset
0.5 % of full scale
Ambient temperature error
within range +10 to +50 C
(compensated temperature range)
zero: 0.02 %/°C typical,
0.05 %/°C max.
span: 0.02 %/°C typical,
0.05 %/°C max.
Characteristic
standard: linear
on request: square root
Deviation from characteristic
1 % of full scale
(limit point adjustment, incl. hysteresis and
repeatability)
Response time
20 msec
Stability per year
0.5 % of full scale
Supply
see ordering details
Ripple
max. 1 V
pp
Max. current drawn
approx. 25 mA
Supply voltage error
0.02 % per V
(with rated supply voltage DC 24 V)
Permissible ambient temperature
-10 to +50 C
Storage temperature
-10 to +70 C
Permissible temperature of medium
-10 to +50 C
Electromagnetic compatibility
to basic standards EN 50081-1 and
EN 50082-1
Mechanical shock
10 g/0.1 msec
Mechanical vibration
5 g max. at 15 to 2000 Hz
Protection
IP65 to EN 60529
(connecting cable dia. 5 mm min.,
7 mm max.)
Housing
ABS
Process connection
see ordering details
Electrical connection
cable gland M12 × 1.5;
internal screw terminals,
for conductor cross-sections up to 1.5 mm
2
Limit contact
• contact rating: 6 A, AC 230 V
• adjustability: 0 to 100 % of full scale
• differential: adjustable,
1to99% of set limit
• output per limit contact:
1 relay changeover contact (floating)
Nominal position
standard: vertical
on request: horizontal
please specify when ordering
Weight
300 g approx.
(400 g approx. with AC supply)
V1.00/EN/00372941
General application
Pressure and differential pressure transmitters Type 4304 are suitable for measuring overpres-
sure, vacuum and differential pressure of non-corrosive gases.
Preferred areas of application are: HVAC, level measurement (bubbling-through method), flow
measurement and monitoring, and filter technology.
On request, the instrument can be supplied with LC display or with LC display and limit con-
tact(s).
Pressure and differential pressure transmitters
Technical data