Agilent Technologies
Agilent 500 Ion Trap LC/MS with the
Agilent 1200 Infinity Series
Quick Start Guide
Step 1. Set up and learn about the system 3
Step 2. Create method and Sequence Sample List 4
Step 3. Start the Sample List in MS Workstation and Sequence in
ChemStation 5
Troubleshooting 10
Use this guide to learn how to set up the Agilent 1200 Infinity Series to
automatically inject samples for analysis on Agilent 500 Ion Trap LC/MS and
MS Workstation software. See Figure 1 for an overview of the workflow.
2 Agilent 1200 Infinity Series LC to Agilent 500 Ion Trap LC/MS Set-Up Guide
Figure 1 Workflow overview
500 Ion Trap LC/MS: How to Acquire MS Data Workflow
In MS Workstation check that 500-MS is
NOT in the AutoStart Module box.
Create MS acquisition method
in MS Workstation.
Create HPLC method in
ChemStation.
Make sure runtimes in MS Workstation and ChemStation are synchronized.
(Allow extra time in LC method for Post Run Events in MS Workstation method.)
Make sure number of vials/injections are synchronized.
Create Sample List in MS
Workstation.
Create Sequence Table
in ChemStation.
m
p
l
e
ti i MS Wktt
ti d
Ch St ti
h
i
Begin MS Workstation
Sample List.
Start Sequence in ChemStation.
S
W
o
Injection and MS analysis of samples begins
que