Solutions for Your Analytical Business
Markets and Applications Programs
1
Abstract
A retention time locked (RTL) method for the determination of trace level volatile organic
compounds (VOCs) in environmental waters was developed on the Agilent 7697A
Headspace and 7890B/5977A GC/MS. Analysis on the GC/MS was performed in
simultaneous Scan/SIM mode. This method can determine 66 VOCs over a calibration
range of 1 – 20 µg/L with some compounds being detectable at lower levels.
This method is therefore suitable for routine analysis, as part of a monitoring strategy
that includes volatile organic compounds, as required by the Water Framework Directive
(2000/60/EC) and detailed in the update to Annex I: Environmental Quality Standards for
Priority Substances and certain other Pollutants [1].
ENVIRONMENTAL ANALYSIS
ANALYSIS OF VOLATILE ORGANIC COMPOUNDS IN
ENVIRONMENTAL WATERS USING THE AGILENT 7697A
HEADSPACE AND 7890B/5977A GC/MS
Solution Note
Author
Bernhard Rothweiler
Agilent Technologies
Waldbronn, Germany
Introduction
The analysis of volatile organic compounds (VOCs) in environmental water samples is
usually performed by either headspace (HS) or purge and trap (P&T), with separation
by gas chromatography (GC) and detection by mass spectrometry (MS).
The P&T technique uses a continuous fl ow of carrier gas to purge any volatiles from the
matrix onto an adsorbant trap. The trap is then heated, releasing the adsorbed compounds
onto the GC/MS for analysis. This technique offers very good sensitivity as the sample
is exhaustively extracted. However, P&T systems are more complicated to run and
maintain compared to HS systems. They can also suffer from water
carryover problems which can lead to reduced sensitivity, loss of peak
shape and cross-contamination. Environmental samples containing
detergents can also suffer from foam formation issues.
2
The other approach is the HS technique, which uses a closed sample container and a sampling system. The water sample in the sealed vial is
heated and agitated to help drive the volatiles from the sample matrix and into the gaseous headspace. An equilibrium is formed between the
two phases. This equilibrium can be shifted by the addition of salt to the sample. After a specified time, a portion of the headspace is transferred
onto the GC/MS via a valve with a sample loop. This technique is robust and experiences few carryover problems as less water is transferred to
the GC/MS.
This application note details the use of an Agilent 7697A HS coupled to a 7890B/5977A GC/MS for the analysis of VOCs in environmental
waters. The MS system features a new inert source with an extractor lens, which provides additional focus to the ion beam into the mass
analyser, resulting in a significant increase in the number of ions analysed and better sensitivity of the instrument. The GC/MS was run in
Scan/SIM mode and the data analysis was accomplished using MassHunter software.
Analytical Technique
Sample Preparation
Sample Preparation: 10 mL of sample and 2 g of salt were added
to each vial. An aliquot of Stock Internal Standard (prepared in
Methanol) was spiked into the solution and the vial sealed.
Standard Preparation: 10 mL water and 2 g of salt were added to
each vial. An aliquot of Stock Standard (prepared in Methanol) and
Stock Internal Standard (prepared in Methanol) were spiked into the
solution and the vial sealed. Standards were prepared at 1, 5, 10 and
20 µg/L, with 66 analytes and 3 internal standards at each level. Two
sets of standards were prepared and run at the beginning and end of
the sequence.
The Internal Standards used were D4- 1,2-Dichloroethane
(CAS 17060-07-0), D8-Toluene (CAS 2037-26-5) and
4-Bromofluorobenzene (CAS 460-00-4). The method was retention
time locked to D8-Toluene at 11.890 minutes.
Instrumentation
Agilent 7697A Headspace Sampler, Agilent 7890B Gas
Chromatograph with an Agilent VF-624 MS 60m capillary column
and an Agilent 5977A Mass Selective Detector with MassHunter
workstation.
Results and Discussion
The method was developed using the 10 µg/L standard solution.
The method is run in Scan/SIM mode, where selective ion
monitoring (SIM) is used for quantitation and Scan could be used for
identification purposes if necessary. To maximise sensitivity the run
has been divided into 16 SIM groups, with 2 ions for each compound,
one for quantitation, the other for qualitation. Figure 1 shows the
Scan (Total Ion Chromatogram, TIC) and SIM traces for the 10 µg/L
standard and Figure 2 the Scan and SIM traces for the 1 µg/L
standard.
Calibration standards at 20, 10, 5 and 1 µg/L were analysed at the
beginning and end of each sequence. Figures 3 and 4 show typical
8-point calibrations for bromochloromethane and 1,2-Dichloroethane.
All VOCs could be detected in the 1 µg/L standard. Some compounds
were even visible in the blanks, mostly at very low levels, indicating
that the detection limits for some of the VOCs may even be below the
0.1 µg/L level.
A spiked surface water sample at the 1 µg/L level was prepared and
analysed to assess reproducibility. Table 1 shows the relative standard
deviation data for this solution (total number of samples = 5). With
the exception of Vinyl chloride, all of the compounds tested for had
RSDs of less than 3% at this level.