V1.00/EN/00576949
Page 1/12Data Sheet 707050
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Email: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 62 50 29
Email: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: +1 315 437 5866
Fax: +1 315 437 5860
Email: info.us@jumo.net
Internet: www.jumousa.com
70705000T10Z001K000
JUMO dTRANS T05
Programmable two-wire transmitter
for installation into terminal head form B and for installation on DIN rail
Brief description
The transmitters record sensor signals from RTD temperature probes, thermocouples, resis-
tance transmitters, or resistors/potentiometers. When using a resistor/potentiometer or RTD tem-
perature probe, the sensor connection on the input side can be connected with a two-wire, three-
wire, or four-wire circuit. Voltage signals in the range from -100 to +1100 mV can also be record-
ed. Depending on the selected measurement input, the linear and temperature-linear lineariza-
tion variants and the possibility of easily configurable customer-specific linearization are
available.
Type 707050 delivers 4 to 20 mA as an output signal. Type 707051 delivers 4 to 20 mA or
0 to 10 V as an output signal. The measurement input and the output signal are galvanically iso-
lated from one another. The output signal can be reversed for both types.
The transmitter configuration with respect to probe type, connection technology of the probe,
measuring range (user configurable), and linearization is carried out by a setup program on the
PC. The connection to the PC is established via a USB interface which does not require addi-
tional auxiliary voltage. Via the USB interface, the min./max. process value and the min./max.
operating temperature recorded by the transmitter can be read and the sensor wiring can be
checked online.
The operating status of the transmitter is indicated by a two-color control LED (red/green). The
control LED is lit green during malfunction-free operation. A fault such as a probe break will be
shown by the corresponding LED signaling.
Type 707050 (dTRANS T05 B)
Type 707051 (dTRANS T05 T)
Block diagram
Transmitter
707050: 4 to 20 mA
707051: 4 to 20 mA
0 to 10 V
Output signals
Voltage supply
Voltage
+
-
RTD
temperature probe
Resistance
transmitter
Input signals
Thermocouple
DC 11 to 35 V
Resistor/
potentiometer
USB interface
Setup
707050: Mini-B
707051: Micro-B
Special features
Measuring input for RTD temperature
probe, thermocouple, resistor/
potentiometer, resistance transmitter,
and voltage
Input and output are electrically isolated
Control LED (red/green)
Configuration directly via USB cable with-
out additional auxiliary voltage
Customer-specific linearization
Detection of the min./max. process value
(drag pointer function including point in
time)
Option to specify the temperature in °F
for temperature sensors
Type 707051 available with screw termi-
nals or spring-cage terminals
V1.00/EN/00576949
Page 2/12Data Sheet 707050
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Email: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 62 50 29
Email: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: +1 315 437 5866
Fax: +1 315 437 5860
Email: info.us@jumo.net
Internet: www.jumousa.com
70705000T10Z001K000
Technical data
Analog input
All analog inputs are equipped with a digital filter of second order (filter constant adjustable from 0 to 10 s) and have a sampling rate of > 2
measurements per second.
RTD temperature probes
Description Standard ITS Connection type Measuring range
in °C
Measuring
accuracy
a
Min. Max.
Pt100 IEC 60751:2008 ITS-90 Two/three-wire -100 200 ±0.2 K
Pt500 Two/three-wire -200 850 ±0.4 K
Pt1000 Four-wire -100 200 ±0.1 K
T
K
=3.85×10
-3
1/K Four-wire -200 850 ±0.2 K
Pt100 GOST 6651-2009 A.2 ITS-90 Two/three-wire -100 200 ±0.2 K
T
K
=3.917×10
-3
1/K Two/three-wire -200 850 ±0.4 K
Four-wire -100 200 ±0.15 K
Four-wire -200 850 ±0.25 K
Pt50 Two/three-wire -200 850 ±0.5 K
T
K
=3.91×10
-3
1/K Four-wire -200 850 ±0.3 K
Ni100 DIN 43760 IPTS-68 Two/three-wire -60 250 ±0.4 K
T
K
=6.18×10
-3
1/K Four-wire -60 250 ±0.2 K
Ni500 Two/three-wire -60 250 ±0.4 K
T
K
=6.18 ×10
-3
1/K Four-wire -60 250 ±0.2 K
Ni1000 Two/three-wire -60 250 ±0.4 K
T
K
=6.18×10
-3
1/K Four-wire -60 250 ±0.2 K
Ni100 GOST 6651-2009 A.5 ITS-90 Two/three-wire -60 180 ±0.4 K
T
K
=6.17×10
-3
1/K Four-wire -60 180 ±0.2 K
Cu50 GOST 6651-2009 A.3 ITS-90 Two/three-wire -180 200 ±0.5 K
T
K
=4.28×10
-3
1/K Four-wire -180 200 ±0.3 K
Cu100 Two/three-wire -180 200 ±0.4 K
T
K
=4.28×10
-3
1/K Four-wire -180 200 ±0.2 K
a
The accuracy values refer to the complete measuring range.
Connection type Two-wire, three-wire, or four-wire circuit
Sensor line resistance
for three and four-wire circuit 11 per wire
for two-wire circuit Measuring resistance + 22 internal conductor resistance
Sensor current < 0.3 mA