P
a
g
e
Find us at www.keysight.com Page 1
Scienlab Battery Test System
Cell Level - Up to 100 A
SL1007A
Battery Test System | Cell Level
Systems up to 6 V | 100 A | 0.6 kW per channel
The Battery Test System (Cell Level) is an electric system which is designed to emulate sink and source for battery
cells for automotive and industrial applications. The bidirectional power supply ensures a very high efficiency.
General system properties
Efficiency
> 90 %
Reactive power
compensation
cos(φ) > 0.98
Mains supply
3 ~, PE, N, 400 VAC (+10 %), 50Hz
System options
Voltage Range
0 to 6 V
Current Range
25 A
50 A
75 A
100 A
Power per Channel
0.15 kW
0.3 kW
0.45 kW
0.6 kW
P
a
g
e
Find us at www.keysight.com Page 2
Measuring specifications
25 A
50 A
75 A
100 A
Current accuracy
± 0.05 %
± 5 mA offset
± 0.05 %
± 10 mA offset
± 0.05 %
± 15 mA offset
± 0.05 %
± 20 mA offset
Voltage accuracy
< 1 mV (typ. 0.5 mV)
Current dynamics
3 ms (10 to 90 % of max. current range) *
Resolution
32 bit
Sampling frequency
1 kHz
Temp. measuring type
1x Pt1000 per channel
Measurement accuracy
± 1 K
Ambient conditions
Storage temperature
5 to 40 °C
Operating temperature
10 to 35 °C
Rel. air humidity
30 to 75 %
Cooling
Cooling type
Water/air heat exchanger
Heat transfer
max. 10 % of total output power
Inlet temperature
10 to 20 °C
Return temperature
max. 30 °C
Connection
¾ “
Documentation
Supplied documents
Operating instructions in English
CE Declaration of Conformity
Acceptance and calibration protocol
System design and realization according to the common basic safety and health
demands of EU directives. Special customer standards are not considered by
default and need an individual discussion and quote
* No switching times at transition from positive to negative current and vice versa