Resistors ; comparison table
Advantages of SSWA (Superior Anti-Surge Wire Wound Axial Resistor)
. Special composite wire-wound structure enhances anti-surge capability. “Patent pending
. Rated to withstand the defibrillation pulse tests described in AAMI EC53/IEC 60601-2-27
. Rated to withstand the ESD pulse tests described in IEC 61000-4-2
(including both the 8KV contact discharge and the 15KV air discharge)
. Best quality and reliability for electrocardiogram(ECG) monitor (3/5/10 leads) applications.
. Reduced total BOM cost with stable production lead time.
SSWA Superior Anti-Surge
Wire Wound Axial Resistors
Brand FIRSTOHM KXX
Construction
Composite wire-wound ceramic resistors
P/N
SSWA01T HXX1 / PXX1
Power Rating
1W 1W
Resistance Value
1K 1K
Body Length
15.5±1.0(mm) 16.0±2.0(mm)
Body Diameter
5.5±0.5(mm) 4.50±1.0(mm)
Tolerance
5% 10%
T.C.R (
PPM/
)
±
200 -900 ~ -1900
Operating TEMP.
-55
~ +200
-40
��
~ +200
Derating Curve
+70
+40
Cost
Low High
SSWA
(Superior Anti-Surge Wire
Wound Axial Resistor)
5KV differential mode
ECG cable Best Solution
ECG cable acts as a necessary route between human body and ECG machine,
there are two reasons to add a resistor inside the wire of ECG cable :
1. To avoid EMI/RFI between each party. 2. For safety to limit any current to
the patient in case of some sort of equipment failure.
The protective component(SSWA) must have sufficient capability to stand
energy of defibrillator pulses but also to stand possible ESD pulses without
specification degradation while maintaining high reliability during the
lifetime of the components.