Specicaons
Safety Standards & EMC Specicaons
Safety Standard Approvals
UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020
TÜV EN 60601-1
EMI Standard EN55011/EN55022, FCC, and VCCI Class B (radiated and conducted)
EMC Performance
EN61000-3-2: Harmonic distoron, Class A and D
EN61000-3-3: Line icker
EN61000-4-2: ESD, ±15 KV air and ± 8KV contact
EN61000-4-3: Radiated immunity, 10V/m
EN61000-4-4: Fast transient/burst, ±2KV
EN61000-4-5: Surge, ±1 KV di., ±2 KV com.
EN61000-4-6: Conducted immunity, 10Vrms
EN61000-4-8: Magnec eld immunity, 30 A/m
EN61000-4-11: Voltage dip immunity, 30% reducon for 500ms,
and 100% reducon for 10ms
*Consult with TT Electronics for informaon on addional country safety approvals
Input Specicaons
Input Voltage Range 90 to 264VAC
Input Frequency Range 47 to 63Hz
Input Current 1.2A (rms) for 115 VAC or 0.6A (rms) for 230 VAC
Earth Leakage Current 180μA max. @ 264VAC, 63Hz
Touch Current 100μA max. @ 264 VAC, 63Hz
Output Specicaons
Ripple & Noise
100 mVp-p maximum on 5V, 6V and 9V outputs, 1% maximum on
other voltage outputs (12V, 13.5V …, 48V etc.)
Overvoltage Protecon Provided and set at 112-140% of nominal output voltage
Overcurrent Protecon Protected to short circuit condions
Temperature Coecient ±0.04%/°C maximum
Transient Response
Maximum excursion of 4% or beer on all models, recovering to 1%
of nal value within 500 us aer a 25% step load change
Environmental Specicaons
Operang Temperature 0°C to +60°C (see derang)
Storage Temperature -40°C to +85°C
Relave Humidity 5% to 95% non-condensing
Temperature Derang Derate from 100% at +40°C, Linearly to 50% at +60°C
General Specicaons
Switching Frequency 50-110KHz
Power Factor 0.98 typical
Eciency
85% min. (except 77% min. for PMP90-10, PMP90-10-1, PMP90-11
and PMP90-12)
Hold-up Time 15ms minimum at 110 VAC
Line Regulaon ±0.5% maximum at full load
Inrush Current 50A @ 115VAC or 100A @ 230VAC, at 25°C cold start
Withstand Voltage
5600VDC from input to output (2 MOPP)
2100VDC from input to ground (1 MOPP)
700VDC from output to ground
(To verify AC strength, get correct test method to avoid power sup-
ply damage.) For Class II models, 4000 VAC from input to output)
MTBF
150,000 hours at full load at 25°C ambient, calculated per MIL-
HDBK-217F