© TT Electronics plc
General Note
All data sheets are subject to change without noce.
ISS.1 22/07/2020 Page: 1
TT Electronics | Power Partners, Inc
43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600
Descripon:
The PMP135 series of AC/DC switching power supplies are for 120-135 was of connuous output power. They are enclosed
in a 94V-0 rated polyphenylene-oxide case with an IEC320/C14 or IEC320/C18 inlet to mate with interchangeable cord for
world-wide use. All models meet EN55011 and FCC class B emission limits, and are designed for medical applicaons.
Features:
Low Safety ground leakage current
Wide input range 90 to 264 VAC
Power factor corrected
200% peak power capability on models below 26Vdc output
Oponal output connectors
Overvoltage protecon
Over temperature protecon
�� Overcurrent protecon
Compliant with CEC and Energy Star Eciency level V requirements
* No Load power consumpon less than 0.5W
* Average acve eciency greater than 87%
Compliant with RoHS requirements
IEC 60601-1-2-4th Edion EMC Compliant
Model
1
Average Acve
Eciency
(typical)
@ 115/230 Vac
Output
Class I Class II V1
Minimum
Current
Maximum
Current
Peak
Current
2
Tol
Ripple &
Noise
3
Max
Power
PMP135-12 PMP135F-12 12V 0A 10A 20A ±5% 120mV 120W 87/89%
PMP135-12-1 PMP135F-12-1 13V 0A 9.23A 18.5A ±5% 130mV 120W 87/89%
PMP135-13 PMP135F-13 14-16V 0A 9.29A 18.6A ±5% 150mV 130W 87/89%
PMP135-13-1 PMP135F-13-1 18-19V 0A 7.50A 15A ±5% 180mV 135W 87/89%
PMP135-13-3 PMP135F-13-3 20-21V 0A 6.75A 13.5A ±5% 200mV 135W 87/89%
PMP135-14 PMP135F-14 24-25V 0A 5.63A 11.3A ±5% 240mV 135W 88/90%
PMP135-15 PMP135F-15 28-29V 0A 4.83A 5.8A ±5% 280mV 135W 88/90%
PMP135-16 PMP135F-16 30-32V 0A 4.50A 5.4A ±5% 300mV 135W 89/90%
PMP135-17 PMP135F-17 36-38V 0A 3.75A 4.5A ±5% 360mV 135W 89/91%
PMP135-18 PMP135F-18 46-50V 0A 2.94A 3.5A ±5% 480mV 135W 90/91%
NOTES:
1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet.
2. For 10 seconds maximum, average power not to exceed maximum power rang.
3. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across
the output.
© TT Electronics plc
General Note
All data sheets are subject to change without noce.
ISS.1 22/07/2020 Page: 2
TT Electronics | Power Partners, Inc
43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600
Specicaons
Safety Standards & EMC Specicaons
Safety Standard Approvals
UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020
TÜV EN 60601-1
EMI Standard EN55011, FCC & VCCI Class B conducted and radiated
EMC Performance
EN61000-3-2: Harmonic distoron, class A and D
EN61000-3-3: Line icker
EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
EN61000-4-3: Radiated immunity, 10 V/m
EN61000-4-4: Fast transient/burst, ±2 KV
EN61000-4-5: Surge, ±1 KV di., ±2 KV com
EN61000-4-6: Conducted immunity, 10 Vrms
EN61000-4-8: Magnec eld immunity, 30 A/m
EN61000-4-11: Voltage dip immunity, 30% reducon for 500 ms,
100% reducon for 10 ms
*Consult with TT Electronics for informaon on addional country safety approvals
Input Specicaons
Input Voltage Range 90 to 264 Vdc
Input Frequency Range 47 to 63Hz
Input Current
1.60 A (rms) for 115 VAC or 0.80 A (rms) for 230 VAC
Earth Leakage Current
200 μA max. @ 264 VAC, 63 Hz
Touch Current
100 μA max. @ 264 VAC, 63 Hz
Power Factor >0.9
Output Specicaons
Ripple and noise 1% peak to peak maximum at the full load
Over Voltage Protecon Provided and set at 115-140% of its nominal output voltage
Over Current Protecon Protected to short circuit condions
Temperature Coecient ±0.04%/°C maximum
Transient Response
Maximum excursion of 4% or beer on all models, recovering to 1%
of nal value within 500 us aer a 25% step load change
Environmental Specicaons
Operang Temperature 0°C to +60°C (See Derang)
Storage Temperature -40°C to +85°C
Relave Humidity 5% to 95% non-condensing
Derang Derate from 100% at +40°C, linearly to 50% at +60°C
General Specicaons
Switching Frequency 90-160KHz
Eciency 87% min. at full load
Hold-up Time 15ms minimum at 110 VAC
Power Factor 0.98 Typical at 115 VAC
Line Regulaon ±0.5% maximum at full load
Inrush Current 80A @ 115 VAC or 160A @ 230 VAC, at 25°C cold start
Withstand Voltage
5600VDC from input to output (2 MOPP)
2100VDC from input to ground (1 MOPP)
700VDC from output to ground
(To verify AC strength, get correct test method to avoid power supply
damage.)
For Class II models, 4000VAC from input to output
MTBF
150,000 hours at full load at 25°C ambient , calculated per MIL-HDBK-
217F