© TT Electronics plc
General Note
All data sheets are subject to change without noce.
ISS.1 22/07/2020 Page: 2
TT Electronics | Power Partners, Inc
43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600
Specicaons
Safety Standards & EMC Specicaons
Safety Standard Approvals
UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020
TÜV EN 60601-1
EMI Standard EN55011, FCC & VCCI Class B conducted and radiated
EMC Performance
EN61000-3-2: Harmonic distoron, class A and D
EN61000-3-3: Line icker
EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
EN61000-4-3: Radiated immunity, 10 V/m
EN61000-4-4: Fast transient/burst, ±2 KV
EN61000-4-5: Surge, ±1 KV di., ±2 KV com
EN61000-4-6: Conducted immunity, 10 Vrms
EN61000-4-8: Magnec eld immunity, 30 A/m
EN61000-4-11: Voltage dip immunity, 30% reducon for 500 ms,
100% reducon for 10 ms
*Consult with TT Electronics for informaon on addional country safety approvals
Input Specicaons
Input Voltage Range 90 to 264 Vdc
Input Frequency Range 47 to 63Hz
Input Current
1.60 A (rms) for 115 VAC or 0.80 A (rms) for 230 VAC
Earth Leakage Current
200 μA max. @ 264 VAC, 63 Hz
Touch Current
100 μA max. @ 264 VAC, 63 Hz
Power Factor >0.9
Output Specicaons
Ripple and noise 1% peak to peak maximum at the full load
Over Voltage Protecon Provided and set at 115-140% of its nominal output voltage
Over Current Protecon Protected to short circuit condions
Temperature Coecient ±0.04%/°C maximum
Transient Response
Maximum excursion of 4% or beer on all models, recovering to 1%
of nal value within 500 us aer a 25% step load change
Environmental Specicaons
Operang Temperature 0°C to +60°C (See Derang)
Storage Temperature -40°C to +85°C
Relave Humidity 5% to 95% non-condensing
Derang Derate from 100% at +40°C, linearly to 50% at +60°C
General Specicaons
Switching Frequency 90-160KHz
Eciency 87% min. at full load
Hold-up Time 15ms minimum at 110 VAC
Power Factor 0.98 Typical at 115 VAC
Line Regulaon ±0.5% maximum at full load
Inrush Current 80A @ 115 VAC or 160A @ 230 VAC, at 25°C cold start
Withstand Voltage
5600VDC from input to output (2 MOPP)
2100VDC from input to ground (1 MOPP)
700VDC from output to ground
(To verify AC strength, get correct test method to avoid power supply
damage.)
For Class II models, 4000VAC from input to output
MTBF
150,000 hours at full load at 25°C ambient , calculated per MIL-HDBK-
217F