Specicaons
Safety Standards & EMC Specicaons
Safety Standard Approvals
UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020
TÜV EN 60601-1
CSA C22.2 No. 62368-1, CSA C22.2 No. 62368-1
TÜV EN 62368-1
EMI Standard EN55011, FCC and VCCI Class B Conducted, class B radiated
EMC Performance
EN61000-3-2: Harmonic Distoron, class A and D
EN61000-3-3: Line Flicker
EN61000-4-2: ESDN, ± 15KV air and ±8 KV contact
EN61000-4-3: Radiated immunity, 10 V/m
EN61000-4-4: Fast transient/ burst, ± 2 KV
EN61000-4-5: Surge, ±1 KV di, ± 2KV com
EN61000-4-6: Conducted Immunity, 10 Vms
EN61000-4-8: Magnec Field immunity, 30 A/m
EN61000-4-11: Voltage dip immunity, 30% reducon for 500 ms, 1005
reducon for 10 ms
*Consult with TT Electronics for informaon on addional country safety approvals
Input Specicaons
Input Voltage Range 90-264 VAC
Input Frequency Range 47-63Hz
Input Current 8.4 A (rms) @115 VAC, 60 Hz 4.2 A (rms) @ 230 VAC, 50 Hz
Earth leakage current 350 µA max. @ 264 VAC, 63 Hz
Touch current 100 µA max. @ 264 VAC, 63 Hz
Output Specicaons
Ripple and noise 1% peak to peak maximum
Remote Sense Compensaon for cable losses up to 0.5V
Overvoltage Protecon
Set at 115-140% of nominal output voltage , latching by recycle input to
reset
Short circuit protecon Automac recovery
Over temperature protecon Latching by recycle input to reset
Thermal Shutdown Protected to over temperature condions
Temperature Coecient All outputs ±0.04% /°C maximum
Transient Response
Maximum excursion of 4%, recovering to 1% of nal value within 500 us
aer a 25% step load change
Standby power 5V at 200mA maximum
Fan Power 12 V at 500 mA maximum
Environmental Specicaons
Operang Temperature -10 ° C to +70°C
Storage Temperature -40 ° C to + 85° C
Relave Humidity 5% to 95% condensing
Temperature Derang
Derate from 100% at +50°C linearly to 50% at +70℃, applicable to convec-
on and forced-air cooling condions
General Specicaons
Switching Frequency 55-300KHz
Power Factor >0.9
Eciency Typical 90%
Hold-up me 20 ms minimum at 100 VAC & 650 W
Line regulaon ±0.5% maximum at full load
Inrush current 20A @ 115VAC, or 40A @ 230VAC, at 25°C cold start
Withstand Voltage
4000 VDC from input to output (2 MOPP)
1500 VDC from input to ground (1 MOPP)
1500 VDC from output to ground
MTBF 190,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F