Revised 8/5/2019
PML065
WWW.PHIHONG.COM
65W 2”x4” Open Frame
Medical Power Supply Series
Features
2x4
Low Cost
Low Profile
Class B EMI
OVP, SCP and OCP
< 100μA Touch Current
Meets DOE Level VI
Class II Design
Applications
Medical
Compliance Approvals
Meets IEC/EN/ANSI/AAMI ES 60601-
1 (ED 3.1)
2X MOPP
IEC60601-1-2(4
th
Ed)
Mechanical Characteristics
Length: 101.6mm (4in)
Height: 28.1mm (1.12in)
Width: 50.8mm (2in)
Weight: 141g (4.97oz)
Output Specifications
Model
Output
Voltage
Output
Power
Regulation
Ripple & Noise
1
P-P(max)
PML065-120
12V
60W
±5%
150mV
PML065-150
15V
65W
±5%
150mV
PML065-180
18V
65W
±5%
180mV
PML065-190
19V
65W
±5%
190mV
PML065-240
24V
65W
±5%
240mV
PML065-480
48V
65W
±5%
480mV
Notes: 1. 20 MHz bandwidth frequency oscilloscope and add a 0.1 μF multilayer cap.and a Low ESR electrolytic cap. (10 μF ) at output
connector terminals. (at nominal line voltage, full load)
Phihong is not responsible for any errors and reserves the right to make changes without notice. Please visit our website at www.phihong.com/
for the most up-to-date specifications and contact information.
INPUT:
AC Input Voltage Range
100 to 240VAC
AC Input Current
1.6A max
Leakage Current
<0.1mA
AC Input Frequency
50-60Hz
AC Inrush Current
80A max/230VAC
(cold start at 25°C, full load)
Input Power Saving
≤0.21W at 230VAC, no load
OUTPUT:
Efficiency
≥88% average at 115VAC & 230VAC
Over Voltage Protection
V out 150%
Short Circuit Protection
Auto-recovery after short-circuit fault being
removed
Over Current Protection
I out 180%
ENVIRONMENTAL:
Temperature
Operation 0 to +50°C
Non-operation -20 to +80°C
Humidity
Operating 20% to 80%
Storage 10% to 90%
EMC
Complies with FCC Class B
Complies with EN55032 Class B
Dielectric Strength
Cut off current 10mA
Primary to secondary 4000VAC for 1min
Insulation Resistance
Primary to secondary 10M ohm for 500VDC
Input Connector
CN1 2PIN WAFER (Pitch=7.92mm)
Output Connector
CN2 4PIN WAFER (Pitch=3.96mm)
PML065 Characteristics