OVERVIEW AND SAFETY PRECAUTIONS
The P200 family of passive probes and accessories provide accurate and safe attachment
to AC and DC test circuits for measurement of signals on analog and digital oscilloscopes in
Category Ill and IV circuits. This probe and accessories adhere
to IEC 61010-031 edition 2015.
PASSIVE PROBE
MARKINGS
WARNING
Failure to follow all instructions in this manual may defeat the safety features and result in
injury or death.
Before using the P200, read and understand the safety warnings, ratings and instructions
contained in this manual.
The P200 family of probes and accessories are not intended for use in wet or damp
locations. Electric shock may result.
There are no user serviceable parts in the P200 probe or accessories.
Do not attempt to open or disassemble.
Probe must be fully connected to the measurement instrument prior to connection to the
device under test.
Probe must be disconnected from the device under test prior to disconnecting from the
measurement instrument.
SPECIFICATIONS, PERFORMANCE AND COMPLIANCE
www.CarlisleIT.com www.CarlisleIT.com
Technical Specifications
Dividing ratio 10:1
Input capacitance 15 pF +/- 2 pF
Compression range 10 pF ~ 25 pF
Frequency range DC-200 MHz
Rise time < 2.1 ns
Cable length 1.4 m
Input impedance 10M ohms ± 1%
Rated voltage when using touch-protected test equipment [a]
[a] Frequency-dependent see derating curve
PB-Free
Processing
Compliant
*EU REACH
Compliant
**EU RoHS2
Compliant
***China
RoHS
Compliant
WEEE
Compliant
ELV
Compliant
Passive Probe
P200 User Manual Guide
Passive Probe
P200 User Manual Guide
Failure to follow all of the instructions
in this manual may defeat the safety
features and result in injury or death.
Double Insulated
UG-10172-011219
SPECIFICATIONS, PERFORMACE AND COMPLIANCE ACCESSORIES
SAFETY HANDLING AND OPERATION
Probe must be fully connected to the measurement instrument prior to connection to
the device under test Probe must be disconnected from the device under test prior to
disconnecting from the measurement instrument.
Barriers are provided to indicate hazardous areas that cannot be touched during operation.
Voltage Ratings
400 V Maximum input voltage of 400 V AC+ DC
300 V Maximum input voltage of 300 V AC+ DC
Category Ratings
CAT II Suitable for connection to category two circuits.
CAT III Suitable for connection to category three circuits.
CAT IV These probes are not suitable for attachment to CAT IV circuits.
Accessories Voltage Ratings
Alligator Ground Lead 400 V CAT II, 300 V CAT III
Hook Tip 400 V CAT II, 300 V CAT III
BNC Adapter Not Applicable
Rf Ground Not Applicable
Screwdriver Not Applicable
Tip Cover Not Applicable
When using the Hook Tip it is safe to use the rib provided for actuation of the hook tip.
CLEANING
When necessary, the probe and accessories can be cleaned with a damp cloth using water
and a mild detergent.
Adequate drying time should be allowed before use.
WARNING
Dry the probe completely before attempting to make voltage measurements.
Screwdriver
Alligator Ground
RF Ground
Tip Cover
Hook tip
BNC Adapter
Passive Probe
P200 User Manual Guide
Passive Probe
P200 User Manual Guide