560550-0078 Issue D Page 1 of 7
GEMS SENSORS & CONTROLS
OPERATING & INSTALLATION INSTRUCTIONS
CE MARKED SERIES 6700B
PLEASE READ CAREFULLY BEFORE INSTALLING
PART NUMBER: 560550-0078 ISSUE: D
Please Note: The Series 6700B is designed and manufactured in accordance with sound
engineering practice as required by the Pressure Equipment Directive 2014/68/EU. This product
must not be used as a “safety accessory” as defined by the Pressure Equipment Directive, Article
1, Point 4. The CE Mark on the unit does not relate to the Pressure Equipment Directive.
INTRODUCTION
Series 6700B are fluid pressure measuring transmitters in which a four active arm Wheatstone
bridge of thin film CVD gauges, integral with a beam structure connected to a pressure summing
diaphragm, is used to convert fluid pressures into a proportional electrical signal.
Common integral electronics converts the input from the gauge to a 4-20mA 2-wire output signal.
Series 6700B conform with the essential protection requirements of the EMC Directive
2004/108/EC amended by certified type testing to EN 61326-1 and EN 61326-2-3.
Conformity with the requirements of the CE mark only applies when the installation
conditions described in these instructions have been met. For units supplied without a
cable assembly connection to the transmitter must be accomplished using Gems Sensors
approved cable. see APPROVED CABLE section.
All instruments conform to the appropriate specifications and/or drawings applicable and have
been subjected to relevant strict quality control procedures.
HAZARDOUS PRODUCTS
The Consumer Protection Act of 1987, Section 6 of the Health and Safety at Work Act 1974 and
the Control of Substances Hazardous to Health Regulations 1988 require that we advise
recipients and users of our products of any potential hazards associated with their storage,
handling or use.
The products which our Company supplies may be classified as Electrical, Electro-Mechanical
and Electronic equipment.
These products are tested and supplied in accordance with our published specifications or
individual special requirements that are agreed in writing at time of order. They are constructed
so as not to affect adversely the safety of persons and property when properly installed,
maintained and used by qualified personnel, in the applications for which they were designed and
manufactured
560550-0078 Issue D Page 2 of 7
ACTION ON RECEIPT
* Check details on Calibration Certificate agree with data etched on transmitter body.
* Check accessories supplied include:
Calibration Certificate (or QA Certificate)
Seal, Dowty Bonded
Mating Connector (where applicable)
GENERAL
* Transmitter should not be subjected to greater than the maximum allowable pressure as
defined on the transmitter label.
* Transmitter should not be subjected to mechanical impact.
* In the event of fire the end user must ensure that the system pressure is vented to a safe
area.
* The effects of decomposition of unstable fluids should be considered by the user when
placing this device in service.
* The pressure transmitter has no means of draining or venting, this must be performed by
another component in the end users system.
* Pressure range must be compatible with the maximum pressure being measured.
* Pressure media must be compatible with the transmitter wetted parts which are 17-4 PH
stainless steels and Nickel Braze to BS 1845 : NK3/HTN2
* Exposed end of cable must be kept free from moisture.
* Liquid must not be allowed to freeze in the pressure port.
Full specifications for all products available on request from our Service Department.
MECHANICAL INSTALLATION
Pressure Connections: G¼ internal pressure connection to BS2779 as standard. Alternative
fitted as specified at time of order.
Pressure couplings screwed into G¼ pressure ports should have a maximum thread engagement
13 mm and UNDER NO CIRCUMSTANCES be allowed to touch the pressure sensitive
diaphragm. Pressure couplings should be sealed against the outer face at the pressure port entry
using bonded seal washer such as:-
Part Number Description
234646-0002 G¼ bonded seal up to +100°C operation.
Dowty Ref: 400-021-4490-02
Refer to Dowty for external pressure ratings.