Rev D
MSHU100 SERIES
2 Watts
Medical DC/DC Converter
Single and Dual Outputs
7/15/2019
Wall Industries, Inc. • Tel: 603-778-2300 Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
Page 1 of 7
Size: 0.94in x 0.54in x 0.35in (24mm x 13.7mm x 9mm)
Low Cost
RoHS Compliant
Efficiency up to 75%
Low Leakage Current
4000VAC I/O Isolation
Water Washable Process Available
UL/cUL/IEC/EN 60950-1 Safety Approvals
and CE Marking
Single and Dual Outputs
Tape & Reel Package Available
Medical EMC Standard meets 4
th
Edition of EMI EN55011 and EMS
EN60601-1-2
Medical Safety meets 1xMOPP &
2xMOPP per 3
rd
Edition of IEC/EN
60601-1 & ANSI/AAMI ES60601-1
DESCRIPTION
The MSHU100 series power modules are 2W DC/DC converters that are specially designed to provide ultra-high levels of isolation (4000VAC) in a
miniature “gull-wing” SMT package. This series meets IPC/JEDEC J-STD-020C 4
th
& 5
th
chapter in solder-reflow for lead free processes. These
converters operate over input voltage ranges of 4.5~5.5VDC, 10.8~13.2VDC, and 21.6~26.4VDC. This series also has single and dual output
voltages of 5, 12, 15, ±12, and ±15VDC. These converters have both medical and industrial safety approvals and are useful for a variety of
applications including distributed power systems, mixed analog/digital subsystems, portable test equipment, local power networks, and battery-
backed systems.
MODEL SELECTION TABLE
Single Output Models
Model Number
(1)
Input Voltage
Range
Output
Voltage
Output Current
Input Current
Max Load
Regulation
Efficiency
Maximum
Capacitive Load
Output
Power
Min.
Max
No Load
Max. Load
MSHU505N2
5 VDC
(4.5 ~ 5.5 VDC)
5 VDC
8mA
400mA
90mA
606mA
12%
66%
330μF
2W
MSHU512N2
12 VDC
3mA
165mA
600mA
10%
66%
330μF
MSHU515N2
15 VDC
2.5mA
133mA
605mA
10%
66%
330μF
MSHU1205N2
12 VDC
(10.8 ~ 13.2 VDC)
5 VDC 8mA 400mA
40mA
253mA 12% 66% 330μF
2W
MSHU1212N2
12 VDC
3mA
165mA
250mA
10%
66%
330μF
MSHU1215N2
15 VDC
2.5mA
133mA
252mA
10%
66%
330μF
MSHU2405N2
24 VDC
(21.6 ~ 26.4 VDC)
5 VDC
8mA
400mA
30mA
126mA
12%
66%
330μF
2W MSHU2412N2 12 VDC 3mA 165mA 125mA 10% 66% 330μF
MSHU2415N2
15 VDC
2.5mA
133mA
126mA
10%
66%
330μF
MODEL SELECTION TABLE
Dual Output Models
Model Number
(1)
Input Voltage
Range
Output
Voltage
Output Current
Input Current
Max Load
Regulation
Efficiency
Maximum
Capacitive Load
Output
Power
Min.
Max
No Load
Max. Load
MSHU512ND2
5 VDC
(4.5 ~ 5.5 VDC)
±12 VDC
±1.5mA
±83mA
90mA
553mA
10%
72%
100#μF
2W
MSHU515ND2
±15 VDC
±1mA
±66mA
542mA
10%
73%
100#μF
MSHU1212ND2
12 VDC
(10.8 ~ 13.2 VDC)
±12 VDC
±1.5mA
±83mA
40mA
224mA
10%
74%
100#μF
2W
MSHU1215ND2 ±15 VDC ±1mA ±66mA 220mA 10% 75% 100#μF
MSHU2412ND2
24 VDC
(21.6 ~ 26.4 VDC)
±12 VDC
±1.5mA
±83mA
30mA
112mA
10%
74%
100#μF
2W
MSHU2415ND2
±15 VDC
±1mA
±66mA
110mA
10%
75%
100#μF
Rev D
MSHU100 SERIES
2 Watts
Medical DC/DC Converter
Single and Dual Outputs
7/15/2019
Wall Industries, Inc. • Tel: 603-778-2300 Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
Page 2 of 7
SPECIFICATIONS
All specifications are based on 25°C, Resistive Load, Nominal Input Voltage, and Rated Output Current unless otherwise noted.
W e reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Input Voltage Range
5V nominal input models
4.5
5
5.5
VDC
12V nominal input models
10.8
12
13.2
24V nominal input models
21.6
24
26.4
Input Surge Voltage (100ms. Max)
5V nominal input models
-0.7
9
VDC
12V nominal input models
-0.7
18
24V nominal input models
-0.7
30
Input Filter
Internal Capacitor
Leakage Current
240VAC, 60Hz
2
μA
OUTPUT SPECIFICATIONS
Output Voltage
See Table
Voltage Accuracy
±2.0
±4.0
%Vnom
Line Regulation
Vin = min. to max.
±1.2
±1.5
%
Load Regulation
Io = 20% to 100%
See Rating Chart
Voltage Balance
Dual Output, Balanced Loads
±0.1
±1.0
%
Output Power
See Table
Output Current
See Table
Maximum Capacitive Load
See Table
Ripple & Noise
0-20MHz Bandwidth
150
mVp-p
PROTECTION
Short Circuit Protection
0.5 seconds max.
Input Fuse Recommendation
5V nominal input models
1000mA slow-blow type
12V nominal input models
500mA slow-blow type
24V nominal input models
200mA slow-blow type
GENERAL SPECIFICATIONS
Efficiency
See Table
Switching Frequency
50
80
100
KHz
Isolation Voltage Rated
60 seconds
4000
VAC
Isolation Voltage Test
Flash Test f or 1 second
6000
VDC
Isolation Resistance
500VDC
10
Isolation Capacitance
100KHz, 1V
15
20
pF
Internal Power Dissipation
650
mW
ENVIRONMENTAL
Operating Temperature (Ambient)
Natural Convection
-25
+80
°C
Operating Temperature (Case)
+90
°C
Storage Temperature
-50
+125
°C
Humidity
Non-Condensing
95
%
Cooling
Natural convec tion
Lead-Free Reflow Solder Process
IPC/JEDEC J-STD-020D.1
Temperature Coefficient
±0.01
±0.02
%/ºC
Moisture Sensitivity Level
IPC/JEDEC J-STD-020D.1
Level 2
MTBF
MIL-HDBK-217F @ 25°C, Ground Benign
2,000,000
hours
PHYSICAL SPECIFICATIONS
Weight
0.13oz (3.75g)
Dimensions (L x W x H)
0.94 x 0.54 x 0.35 inches
24.0 x 13.7 x 9.0 mm
Case Material
Non-conductive Black Plastic
(Flammability to UL 94V-0 rated)
Pin Material
Phosphor Bronze
Flammability
UL94V-0
SAFETY CHARACTERISTICS
Safety Standards
UL/cUL 60950-1
(5)
, CSA C22.2 No. 60950-1
ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1
IEC/EN 60950-1, IEC/EN 60601-1 3
rd
Edition 1xMOPP & 2xMOPP
Safety Approvals
UL/cUL 60950-1
(5)
Recognition (UL Certificate)
IEC/EN 60950-1 (CB Report)
ANSI/AAMI ES60601-1 1xMOPP & 2xMOOP recognition (UL Certificate
IEC/EN 60601-1 3
rd
Edition (CB Report)