36 Thornwood Dr. – Ithaca, NY 14850
tel: 607-257-1080 – fax:607-257-1146
www.kionix.com - info@kionix.com
Introduction
Kionix calibrates the offset of the KX022 and KX023 at the factory. After
assembly, it is sometimes found by the customer that there is a subsequent
offset shift.
The KX022 and KX023 have embedded registers used to provide a final
factory calibration. These registers have the default factory calibration but
also they can be adjusted by a user programmatically for an additional offset
adjustment to the original factory programmed calibration.
The values for the user applied offset adjustment are stored in RAM and must
be updated by the controlling system after each power up or reset.
Offset Calibration readings and adjustments are calculated using 2g High RES
mode.
The Offset calibration adjustment adds a constant value to correct for an
offset error.
Two registers are employed to make the Offset Calibration adjustment.